平扫多排CT定量肝脏脂肪含量:体模与临床中与MRI质子密度脂肪分数的相关性
研究构建含12个脂质分数0%–100%小瓶的CT-MRI体模,并回顾2007至2017年221例同时行平扫CT与化学位移编码MRI的患者,评估平扫CT肝衰减值与MRI质子密度脂肪分数(PDFF)的关系。体模r²=0.986,临床120kV且1个月内r²=0.828,给出换算公式MRI PDFF(%)=-0.58×CT衰减值(HU)+38.2;间隔超过1个月、非120kV及铁过载时相关性变差。
要点
- 平扫CT肝衰减值与MRI PDFF呈强线性相关,120kV且1个月内r²=0.828。
- 临床换算公式为MRI PDFF(%)=-0.58×CT衰减值(HU)+38.2。
- CT与MRI间隔超过1个月、非120kV或铁过载时相关性变差,CT低估肝脂肪含量。
核心结论
- 平扫CT衰减值与MRI PDFF的线性相关可在临床实践中用于定量肝脏脂肪含量;相关性随CT-MRI时间间隔延长、管电压设置不同及铁过载而变差。
研究设计与样本
- 体模研究加回顾性临床研究:体模含12个脂质分数0%–100%的小瓶;临床纳入221例(男120例、女101例,平均年龄54岁)于2007至2017年行平扫CT和化学位移编码MRI,其中92例有多次120kV CT扫描,120kV且1个月内行CT者72例作为线性相关主要队列,以共配准ROI测量右肝CT衰减值和MRI PDFF。
诊断标准与关键数值
- 体模CT衰减值与MRI PDFF线性拟合r²=0.986;临床120kV且1个月内CT与MRI PDFF线性相关r²=0.828,换算公式为MRI PDFF(%)=-0.58×CT衰减值(HU)+38.2;CT-MRI间隔超过1个月r²=0.565;非120kV设置r²=0.554。
临床应用与阅片要点
- 作者认为平扫CT衰减值与MRI PDFF的线性相关可在临床实践中用于定量肝脏脂肪含量;阅片时需注意CT-MRI间隔超过1个月、非120kV管电压及铁过载会使相关性变差,铁过载患者CT可低估肝脂肪含量。
局限性
- 回顾性临床研究。
- CT与MRI间隔超过1个月时相关性变差(r²=0.565)。
- 该特定关系对非120kV设置适用性较差(r²=0.554)。
- 铁过载患者CT肝脂肪含量被低估。
- 多次扫描患者相关性随时间推移逐渐变差。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study was to evaluate the relation between unenhanced CT liver attenuation values and MRI-derived proton density fat fraction (PDFF) for estimation of liver fat content at CT.
MATERIALS AND METHODS: A CT-MRI phantom was constructed and imaged containing 12 vials with lipid fractions ranging from 0% to 100%. For the retrospective clinical arm, 221 patients (120 men, 101 women; mean age, 54 years) underwent both unenhanced CT and chemical shift-encoded MRI of the liver between 2007 and 2017. Among these patients, 92 had more than one 120-kV CT scan for comparison. CT attenuation and MRI PDFF were derived with coregistered ROI measurements in the right hepatic lobe. The 120-kV subgroup of CT examinations performed within 1 month of MRI PDFF examinations (n = 72) served as the primary cohort for linear correlation. The effects of different tube voltage settings, time intervals between CT and MRI, and iron overload were assessed. Linear least squares regression analysis was performed.
RESULTS: Phantom results showed excellent linear fit between CT attenuation and MRI PDFF (r2 = 0.986). In patients, 120-kV CT performed within 1 month of MRI PDFF exhibited strong linear correlation (r2 = 0.828) that closely matched the phantom data, yielding the following clinical CT-MRI conversion formula: MRI PDFF (%) = -0.58 × CT attenuation (HU) + 38.2. Correlation worsened for CT-to-MRI intervals longer than 1 month (r2 = 0.565), and this specific relationship did not apply as well to non-120-kV settings (r2 = 0.554). For patients with multiple scans, correlation progressively worsened over time. CT-based liver fat content was underestimated in several patients with iron overload.
CONCLUSION: The linear correlation between unenhanced CT attenuation and MRI PDFF allows quantification of liver fat content by means of unenhanced CT in clinical practice. As expected, correlation worsened with increasing CT-MRI time interval, variable tube voltage settings, and iron overload.