在儿童心胸CT管电流调制中,扫描范围调整前还是后输入体型适配的容积CT剂量指数更利于辐射剂量优化
本研究在83例儿童心胸CT管电流调制检查中,比较扫描范围调整后(42例)与调整前(41例)输入体型适配CTDIvol对辐射剂量和图像质量的影响。调整前输入组CT扫描后CTDIvol显著低于调整后输入组(1.4±0.3 mGy vs 1.7±0.1 mGy,p<0.0001),图像噪声和图像质量无显著差异,结论为扫描范围调整前输入CTDIvol可在图像质量相当的情况下显著降低剂量(18%)。
要点
- 儿童心胸CT管电流调制时,扫描范围调整前输入CTDIvol较调整后输入可使CT扫描后CTDIvol降低18%。
- 两组图像噪声和图像质量评分无显著差异(p>0.05)。
- 扫描范围调整后输入组的CT扫描后CTDIvol与图像噪声呈中度正相关(r=0.49,p≤0.001),调整前输入组无显著相关。
核心结论
- 儿童心胸CT管电流调制中,扫描范围调整前输入体型适配CTDIvol较调整后输入可显著降低剂量(18%),图像质量相当。
研究设计与样本
- 83例患儿行儿童心胸CT管电流调制检查,扫描范围调整后输入体型适配CTDIvol组42例(组1),扫描范围调整前输入组41例(组2),比较两组患者相关参数、辐射剂量和图像质量参数。
诊断标准与关键数值
- 组1 CT扫描后CTDIvol为1.7±0.1 mGy,组2为1.4±0.3 mGy,组1显著高于组2(p<0.0001)。
- 组1图像噪声为4.6±0.5 HU,组2为4.5±0.7 HU,两组无显著差异(p>0.05)。
- 组1图像质量为1.5±0.6,组2为1.5±0.6,两组无显著差异(p>0.05)。
- 两组除体密度外,所有患者相关参数与CT扫描前后CTDIvol均呈正相关(r=0.49-0.94;p<0.01)。
- 组1 CT扫描后CTDIvol与图像噪声呈中度正相关(r=0.49;p≤0.001),组2无显著相关(p>0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To determine whether the body size-adapted volume computed tomography (CT) dose index (CTDvol) in pediatric cardiothoracic CT with tube current modulation is better to be entered before or after scan range adjustment for radiation dose optimization.
MATERIALS AND METHODS: In 83 patients, cardiothoracic CT with tube current modulation was performed with the body size-adapted CTDIvol entered after (group 1, n = 42) or before (group 2, n = 41) scan range adjustment. Patient-related, radiation dose, and image quality parameters were compared and correlated between the two groups.
RESULTS: The CTDIvol after the CT scan in group 1 was significantly higher than that in group 2 (1.7 ± 0.1 mGy vs. 1.4 ± 0.3 mGy; p < 0.0001). Image noise (4.6 ± 0.5 Hounsfield units [HU] vs. 4.5 ± 0.7 HU) and image quality (1.5 ± 0.6 vs. 1.5 ± 0.6) showed no significant differences between the two (p > 0.05). In both groups, all patient-related parameters, except body density, showed positive correlations (r = 0.49-0.94; p < 0.01) with the CTDIvol before and after the CT scan. The CTDIvol after CT scan showed modest positive correlation (r = 0.49; p ≤ 0.001) with image noise in group 1 but no significant correlation (p > 0.05) in group 2.
CONCLUSION: In pediatric cardiothoracic CT with tube current modulation, the CTDIvol entered before scan range adjustment provides a significant dose reduction (18%) with comparable image quality compared with that entered after scan range adjustment.