采用分区收发阵列的矩阵经食管超声心动图探头进行快速容积成像
本文描述一种三维多线平行波束合成方案,用于实时容积超声成像,采用对角切割阵元且收发阵列分离的原型矩阵经食管超声心动图探头。通过85个收发事件序列实现60°×60°容积扇区成像,容积率最高达59 Hz,在健康成年猪体内以22 Hz容积率清晰显示心脏结构及瓣膜运动。
要点
- 采用分区收发阵列与微波束合成专用集成电路的矩阵经食管探头,实现实时三维容积成像。
- 85个收发事件序列在60°×60°容积扇区中表现良好,容积率最高达59 Hz。
- 健康成年猪体内以22 Hz容积率清晰显示心脏结构,包括瓣膜运动。
核心结论
- 所提出的平行波束合成方案可实现高达59 Hz的容积率,并通过角度加权组合重叠子容积产生良好质量的图像,体内成像清晰显示心脏结构及瓣膜运动。
研究设计与样本
- 研究描述一种三维多线平行波束合成方案,采用对角切割阵元且收发阵列分离的原型矩阵经食管超声心动图探头,通过Field II仿真、体模测量和体内成像进行验证。
关键影像表现/征象
- 体内三维成像以22 Hz容积率在健康成年猪中清晰显示心脏结构,包括瓣膜运动。
- 体模实验中的点扩散函数、对比度和对比噪声比与仿真结果密切匹配。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
We describe a 3-D multiline parallel beamforming scheme for real-time volumetric ultrasound imaging using a prototype matrix transesophageal echocardiography probe with diagonally diced elements and separated transmit and receive arrays. The elements in the smaller rectangular transmit array are directly wired to the ultrasound system. The elements of the larger square receive aperture are grouped in 4 × 4-element sub-arrays by micro-beamforming in an application-specific integrated circuit. We propose a beamforming sequence with 85 transmit-receive events that exhibits good performance for a volume sector of 60° × 60°. The beamforming is validated using Field II simulations, phantom measurements and in vivo imaging. The proposed parallel beamforming achieves volume rates up to 59 Hz and produces good-quality images by angle-weighted combination of overlapping sub-volumes. Point spread function, contrast ratio and contrast-to-noise ratio in the phantom experiment closely match those of the simulation. In vivo 3-D imaging at 22-Hz volume rate in a healthy adult pig clearly visualized the cardiac structures, including valve motion.
引用
- Bera D, van den Adel F, Radeljic-Jakic N, et al. Fast Volumetric Imaging Using a Matrix Transesophageal Echocardiography Probe with Partitioned Transmit-Receive Array. Ultrasound Med Biol. 2018. PMID: 30037476. doi:10.1016/j.ultrasmedbio.2018.05.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。