文献速递外周血管腹部USMR

腹主动脉高帧率超声造影粒子图像测速的首批人体结果

High-Frame-Rate Contrast-enhanced US Particle Image Velocimetry in the Abdominal Aorta: First Human Results.
PubMed 文献 · Radiology · 2018-07-17

研究纳入15名健康受试者,采用高帧率超声造影粒子图像测速(回声PIV)评估腹主动脉血流,并与相位对比MRI比较。结果显示所有受试者均可成功显示主动脉分叉,最低造影剂剂量0.25 mL联合最低机械指数(0.01或0.03)时回声PIV效果最佳,流速剖面与MRI吻合良好(相似指数0.98和0.99,峰值速度差异8.5%和17.0%)。结论认为回声PIV用于人体腹主动脉血流定量可行,在主动脉疾病临床评估中有潜在应用价值。

要点

  • 15名健康受试者中高帧率超声造影均成功显示腹主动脉分叉。
  • 0.25 mL造影剂联合机械指数0.01或0.03时回声PIV效果最佳。
  • 机械指数≥0.06时出现明显微泡破坏。

核心结论

  • 高帧率超声造影粒子图像测速(回声PIV)用于人体腹主动脉血流定量可行,在主动脉疾病临床评估中有潜在应用价值。

研究设计与样本

  • 可行性研究,纳入15名健康受试者(6名男性;中位年龄23岁,年龄范围18-34岁;中位体重指数20.3 kg/m2,范围17.3-24.9 kg/m2),行高帧率超声造影检查并以相位对比MRI作为参照。

关键影像表现/征象

  • 所有受试者均成功显示腹主动脉分叉。
  • 回声PIV与MRI图像中血流模式定性相似。
  • 回声PIV与MRI流速剖面吻合良好,时间剖面相似指数0.98,空间剖面相似指数0.99。
  • 峰值速度差异在时间剖面为8.5%,在空间剖面为17.0%。

诊断标准与关键数值

  • 造影剂微泡按递增剂量0.25、0.5、0.75和1.5 mL注射,每剂后行超声测量以确定最佳剂量。
  • 评估的机械指数值为0.09、0.06、0.03和0.01,采用发散波采集方案。
  • 最低造影剂剂量0.25 mL联合最低机械指数(0.01或0.03)时获得最佳回声PIV结果。
  • 机械指数≥0.06时发生明显微泡破坏。
  • PIV分析通过对所有采集图像进行两两互相关完成。

临床应用与阅片要点

  • 作者认为回声PIV在主动脉疾病临床评估中有潜在应用价值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To study the feasibility of high-frame-rate (HFR) contrast material-enhanced (CE) ultrasound particle image velocimetry (PIV), or echo PIV, in the abdominal aorta. Materials and Methods Fifteen healthy participants (six men; median age, 23 years [age range, 18-34 years]; median body mass index, 20.3 kg/m2 [range, 17.3-24.9 kg/m2]) underwent HFR CE US. US microbubbles were injected at incremental doses (0.25, 0.5, 0.75, and 1.5 mL), with each dose followed by US measurement to determine the optimal dosage. Different US mechanical index values were evaluated (0.09, 0.06, 0.03, and 0.01) in a diverging wave acquisition scheme. PIV analysis was performed via pairwise cross-correlation of all captured images. Participants also underwent phase-contrast MRI. The echo PIV and phase-contrast MRI velocity profiles were compared via calculation of similarity index and relative difference in peak velocity. Results Visualization of the aortic bifurcation with HFR CE US was successful in all participants. Optimal echo PIV results were achieved with the lowest contrast agent dose of 0.25 mL in combination with the lowest mechanical indexes (0.01 or 0.03). Substantial bubble destruction occurred at higher mechanical indexes (≥0.06). Flow patterns were qualitatively similar in the echo PIV and MR images. The echo PIV and MRI velocity profiles showed good agreement (similarity index, 0.98 and 0.99; difference in peak velocity, 8.5% and 17.0% in temporal and spatial profiles, respectively). Conclusion Quantification of blood flow in the human abdominal aorta with US particle image velocimetry (echo PIV) is feasible. Use of echo PIV has potential in the clinical evaluation of aortic disease. © RSNA, 2018 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Engelhard S, Voorneveld J, Vos HJ, et al. High-Frame-Rate Contrast-enhanced US Particle Image Velocimetry in the Abdominal Aorta: First Human Results. Radiology. 2018. PMID: 30015586. doi:10.1148/radiol.2018172979

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。