文献速递胸部CT

前向投影模型迭代重建在肺癌低剂量筛查CT中的应用:与自适应迭代剂量降低3D的比较

Forward-Projected Model-Based Iterative Reconstruction in Screening Low-Dose Chest CT: Comparison With Adaptive Iterative Dose Reduction 3D.
PubMed 文献 · AJR Am J Roentgenol · 2018-07-24

回顾性分析50例连续患者低剂量胸部CT,比较FIRST与AIDR 3D重建在图像噪声、主观质量及肺结节检出、大小和定性上的差异。FIRST图像噪声更低(肺窗降低44%,软组织窗降低32%),主观质量相当,结节计数、大小和定性一致性良好。

要点

  • FIRST较AIDR 3D肺窗噪声降低44%(41±7 vs 74±8 HU),软组织窗降低32%。
  • 两种重建主观图像质量相当(FIRST肺窗Likert评分3.0±0.3,软组织窗3.2±0.3)。
  • FIRST与AIDR 3D在结节计数、大小及定性上一致性良好(ICC 0.78–0.99,kappa 0.82–0.92)。

核心结论

  • FIRST重建图像在噪声方面优于AIDR 3D,在主观图像质量、肺结节计数和结节定性方面与AIDR 3D等效。

研究设计与样本

  • 回顾性研究,纳入2015年12月至2016年1月间50例连续患者的低剂量胸部CT,分别用FIRST和AIDR 3D重建肺窗和软组织窗图像,由阅片者独立评估。

关键影像表现/征象

  • FIRST肺窗图像噪声41±7 HU,AIDR 3D为74±8 HU,FIRST降低44%。
  • FIRST软组织窗图像噪声11±2 HU,AIDR 3D为16±2 HU,FIRST降低32%。
  • FIRST主观图像质量与AIDR 3D相当:软组织窗重建Likert评分3.2±0.3,肺窗重建3.0±0.3。

诊断标准与关键数值

  • 主观图像质量采用5分Likert量表:1分远优,2分优,3分相当,4分差,5分远差。
  • 结节计数一致性ICC:阅片者1为0.83,阅片者2为0.78。
  • 结节大小一致性ICC:阅片者1为0.99,阅片者2为0.99。
  • 最大结节定性(实性、部分实性、磨玻璃)一致性kappa:阅片者1为0.92,阅片者2为0.82。

临床应用与阅片要点

  • 在低剂量肺癌筛查CT中,FIRST重建可降低图像噪声,同时保持与AIDR 3D相当的主观图像质量和结节评估能力。
  • 阅片时需注意FIRST与AIDR 3D在结节计数、大小和定性上的一致性良好,但结节计数ICC为0.78–0.83,提示存在一定变异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The objective of this study is to compare forward-projected model-based iterative reconstruction solution (FIRST), a newer fully iterative CT reconstruction method, with adaptive iterative dose reduction 3D (AIDR 3D) in low-dose screening CT for lung cancer. Differences in image noise, image quality, and pulmonary nodule detection, size, and characterization were specifically evaluated.

MATERIALS AND METHODS: Low-dose chest CT images obtained for 50 consecutive patients between December 2015 and January 2016 were retrospectively reviewed. Images were reconstructed using FIRST and AIDR 3D for both lung and soft-tissue reconstruction. Images were independently reviewed to assess image noise, subjective image quality (with use of a 5-point Likert scale, with 1 denoting far superior image quality; 2, superior quality; 3, equivalent quality; 4, inferior quality; and 5, far inferior quality), pulmonary nodule count, size of the largest pulmonary nodule, and characterization of the largest pulmonary nodule (i.e., solid, part solid, or ground glass).

RESULTS: Across all 50 cases, measured image noise was lower with FIRST than with AIDR 3D (lung window, 44% reduction, 41 ± 7 vs 74 ± 8 HU, respectively; soft-tissue window, 32% reduction, 11 ± 2 vs 16 ± 2 HU, respectively). Readers subjectively rated images obtained with FIRST as comparable to images obtained with AIDR 3D (mean [± SD] Likert score for FIRST vs AIDR 3D, 3.2 ± 0.3 for soft-tissue reconstructions and 3.0 ± 0.3 for lung reconstructions). For each reader, very good agreement regarding nodule count was noted between FIRST and AIDR 3D (interclass correlation coefficient [ICC], 0.83 for reader 1 and 0.78 for reader 2). Excellent agreement regarding nodule size (ICC, 0.99 for reader 1 and 0.99 for reader 2) and characterization of the largest nodule (kappa value, 0.92 for reader 1 and 0.82 for reader 2) also existed.

CONCLUSION: Images reconstructed with FIRST are superior to those reconstructed AIDR 3D with regard to image noise and are equivalent with regard to subjective image quality, pulmonary nodule count, and nodule characterization.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Hassani C, Ronco A, Prosper AE, et al. Forward-Projected Model-Based Iterative Reconstruction in Screening Low-Dose Chest CT: Comparison With Adaptive Iterative Dose Reduction 3D. AJR Am J Roentgenol. 2018. PMID: 30040468. doi:10.2214/AJR.17.19245

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。