CT引导下机器人插入多种消融针的准确性:动物实验研究
在6头猪中,使用机器人于CT引导下将四种消融针(Cool-tip、LeVeen、IceRod、Emprint)分别向肝、肾、肺、臀肌的1.0 mm靶点各尝试插入5次,共评估插入准确性;总体平均准确性为2.8 mm,不同针型(P=0.38)和不同部位(P=0.53)间均无显著差异。结论认为CT引导下机器人插入多种消融针在猪体内是准确的,与针型和部位无关。
要点
- 四种消融针总体平均插入准确性为2.8 mm。
- 不同针型间插入准确性无显著差异(P=0.38)。
- 不同部位间插入准确性无显著差异(P=0.53)。
核心结论
- CT引导下机器人插入多种消融针在猪体内是准确的,与针型或部位无关。
研究设计与样本
- 动物实验,6头猪,每种消融针向肝、肾、肺、臀肌的1.0 mm靶点各尝试插入5次,共评估四种消融针(Cool-tip、LeVeen、IceRod、Emprint),以靶点中心与针尖之间的三维长度评价插入准确性,采用单因素方差分析比较。
诊断标准与关键数值
- 插入准确性定义为靶点中心与针尖之间的三维长度。
- 所有四种针在所有四个部位的总平均准确性为2.8 mm。
- Cool-tip针、LeVeen针、IceRod针和Emprint针的平均插入准确性分别为2.8 mm、3.1 mm、2.5 mm和2.7 mm。
- 肝、肾、肺和臀肌的平均插入准确性分别为2.7 mm、2.9 mm、2.9 mm和2.5 mm。
- 不同针型间插入准确性无显著差异(P=0.38),不同部位间也无显著差异(P=0.53)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the accuracy of robotic insertion of various ablation needles at various locations under computed tomography (CT) guidance in swine.
MATERIALS AND METHODS: The robot was used for CT-guided insertion of four ablation needles, namely a single internally cooled radiofrequency ablation (RFA) needle (Cool-tip), a multi-tined expandable RFA needle (LeVeen), a cryoablation needle (IceRod), and an internally cooled microwave ablation needle (Emprint). One author remotely operated the robot with the operation interface in order to orient and insert the needles under CT guidance. Five insertions of each type of ablation needle towards 1.0-mm targets in the liver, kidney, lung, and hip muscle were attempted on the plane of an axial CT image in six swine. Accuracy of needle insertion was evaluated as the three-dimensional length between the target centre and needle tip. The accuracy of needle insertion was compared according to the type of needle used and the location using one-way analysis of variance.
RESULTS: The overall mean accuracy of all four needles in all four locations was 2.8 mm. The mean accuracy of insertion of the Cool-tip needle, LeVeen needle, IceRod needle, and Emprint needle was 2.8 mm, 3.1 mm, 2.5 mm, and 2.7 mm, respectively. The mean accuracy of insertion into the liver, kidney, lung, and hip muscle was 2.7 mm, 2.9 mm, 2.9 mm, and 2.5 mm, respectively. There was no significant difference in insertion accuracy among the needles (P = .38) or the locations (P = .53).
CONCLUSION: Robotic insertion of various ablation needles under CT guidance was accurate regardless of type of needle or location in swine.