长治疗体积放疗患者体位俯仰角对剂量的影响及六自由度床的实用性

Dosimetric influence of pitch in patient positioning for radiotherapy of long treatment volumes; the usefulness of six degree of freedom couch.
PubMed 文献 · Br J Radiol · 2018-07-25

纳入10例鼻咽癌和10例食管癌患者,模拟计划CT腹侧和背侧倾斜±1.5°和±3°的俯仰角,计算验证计划并与原计划及危及器官剂量约束比较。靶区剂量偏差随俯仰角增大而增加,鼻咽癌平均变化达2%、食管癌达1%;鼻咽癌患者脑干、脊髓、晶状体、口腔黏膜和腮腺剂量受影响,食管癌无显著变化。结论认为俯仰角误差对鼻咽癌危及器官有临床相关性,推荐俯仰角偏差超过1.5°时使用六自由度床校正。

要点

  • 鼻咽癌患者俯仰角3°背侧倾斜时脊髓最大剂量增加10.0%、脑干最大剂量增加6.0%。
  • 鼻咽癌患者俯仰角3°腹侧倾斜时腮腺平均剂量增加4.3%、晶状体最大剂量增加3.3%、口腔黏膜平均剂量增加2.6%。
  • 食管癌患者各倾斜角度下危及器官剂量均无显著变化,所有计划临床可接受。

核心结论

  • 两种肿瘤的靶区覆盖仅受轻微影响,但俯仰角误差对鼻咽癌患者的危及器官可能具有临床相关性;推荐鼻咽癌患者俯仰角偏差超过1.5°时使用六自由度床校正,以避免危及器官剂量超限。

研究设计与样本

  • 纳入10例鼻咽癌和10例食管癌患者,均接受容积调强弧形治疗;通过将计划CT向腹侧和背侧倾斜±1.5°和±3°模拟俯仰角,在倾斜数据集上计算验证计划并与原计划及危及器官剂量约束比较。

诊断标准与关键数值

  • 鼻咽癌靶区剂量偏差随俯仰角增大而增加,平均变化达2%;食管癌平均变化达1%。
  • 鼻咽癌患者3°背侧倾斜时脑干最大剂量+6.0%、脊髓最大剂量+10.0%;3°腹侧倾斜时晶状体最大剂量+3.3%、口腔黏膜平均剂量+2.6%、腮腺平均剂量+4.3%。
  • 食管癌患者各倾斜角度下危及器官剂量无显著变化。
  • 鼻咽癌计划在1.5°腹侧倾斜时1例、3°腹侧倾斜时2例、3°背侧倾斜时2例共5例计划因危及器官剂量不再临床可接受;食管癌所有倾斜计划均临床可接受。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE:: Pitch, the rotation around the transversal axis of the patient during radiotherapy has little impact on the dose distribution of small spherical treatment volumes; however it might affect treatment of long volumes requiring a correction with a six degree of freedom couch.

METHODS:: We included 10 patients each with nasopharyngeal carcinoma (NPC) and esophageal cancer, treated with volumetric modulated arc therapy. Pitch was simulated by tilting the planning CT in ventral and dorsal direction by ± 1.5° and ± 3°. Verification plans were calculated on the tilted datasets and were compared to the original plan and the dose constraints of the organs at risk (OAR).

RESULTS:: The deviation in dose to the planning target volume is increasing with the degree of pitch with mean changes of up to 2% for NPC and 1% for esophageal cancer. The most affected OAR in NPC patients are brainstem (max. dose +6.0%) and spinal cord (max. dose +10.0%) when tilted by 3° dorsally and lenses (max. dose +3.3%), oral mucosa (mean dose +2.6%) and parotid glands (mean dose +4.3%) when tilted by 3° ventrally. For esophageal cancer patients, there was no significant change in dose to any OAR. Whereas for esophageal cancer, all tilted treatment plans were still clinically acceptable regarding OAR, 5 NPC plans would no longer be acceptable with a pitch of 1.5° ventral (N = 1), 3° ventral (N = 2) and 3° dorsal (N = 2).

CONCLUSION:: Planning target volume coverage in both tumor entities was only slightly affected, but pitch errors could be relevant for OAR in NPC patients.

ADVANCES IN KNOWLEDGE:: A correction with a six degree of freedom couch is recommended for NPC patients with a pitch mismatch of more than 1.5° to avoid exceeded doses to the OAR.

来源:Br J Radiol,查看原文。版权归原作者所有。