使用O-15水PET/MRI与ASL相关性无创测量脑血流的方法
研究纳入11名健康志愿者,采用H215O-PET/MRI无创图像衍生输入函数法测量脑血流,并用颅骨体模进行交叉校准。结果显示ARG法、1-TCM法和ASL-MRI的全局平均脑血流分别为44.8±4.3、47.9±5.9和57.9±8.6 mL/min/100 g,ASL-MRI显著高于H215O-PET(P<0.001),但各方法间显著相关(P<0.05)。结论认为无创IDIF法可提供合理的定量值,ASL-MRI倾向于高估脑血流。
要点
- ASL-MRI全局平均脑血流57.9±8.6 mL/min/100 g,显著高于H215O-PET的44.8±4.3(ARG法)和47.9±5.9(1-TCM法)(P<0.001)。
- 三种方法测得的脑血流值在散点图上彼此显著相关(P<0.05)。
- 颅骨体模交叉校准支持IDIF法可靠估计动脉放射性浓度。
核心结论
- H215O-PET/MRI联合IDIF及颅骨体模交叉校准法可无创测量脑血流并提供合理定量值,IDIF法能可靠估计动脉放射性浓度,ASL-MRI同步采集的灌注图像倾向于高估脑血流。
研究设计与样本
- 研究纳入11名健康志愿者(9名男性,43.9±10.9岁),行3分钟H215O-PET及包括动脉自旋标记(ASL)灌注MRI在内的多次MRI扫描,PET图像采用两组重建参数重建为动态数据,重建参数由体模管放射性浓度可重复性基础评估确定。
诊断标准与关键数值
- ARG法测得的全局平均脑血流为44.8±4.3 mL/min/100 g。
- 1-TCM法测得的全局平均脑血流为47.9±5.9 mL/min/100 g。
- ASL-MRI测得的全局平均脑血流为57.9±8.6 mL/min/100 g,显著高于H215O-PET(P<0.001)。
- 三种方法的脑血流值在散点图上彼此显著相关(P<0.05)。
临床应用与阅片要点
- IDIF法从PET数据的多个图像层面提取颈内动脉的时间-放射性曲线,用于无创估算动脉输入函数。
- ASL-MRI与H215O-PET同步采集时,ASL-MRI测得的脑血流值倾向于高估,解读时应结合方法学差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: A noninvasive image derived input function (IDIF) method was applied to estimate arterial input function from brain H215O-PET/MRI images for the measurement of cerebral blood flow (CBF) because of difficulty in arterial blood sampling during PET/MRI scans. To evaluate accuracy and reproducibility of radioactivity in the internal carotid arteries (ICA) for the IDIF method, a new phantom using a skull bone was applied in the cross-calibration process between the scanner and a gamma-well counter.
METHODS: Eleven healthy volunteers (9 males, 43.9 ± 10.9y) underwent PET/MRI studies with a 3-min H215O-PET and several MRI scans including arterial spin labeling (ASL) perfusion MRI. PET images were reconstructed as dynamic data using two sets of reconstruction parameters, which were determined by basic assessment of radioactivity concentration reproducibility in the tubes of the phantom. The IDIF method extracted the time-activity curves of the ICA from several image slices in the PET data. CBF images were calculated using the autoradiographic (ARG) method and a one-tissue compartment model (1-TCM).
RESULTS: The global means of CBF from the ARG, 1-TCM, and ASL-MRI were 44.8 ± 4.3, 47.9 ± 5.9 and 57.9 ± 8.6 (mL/min/100 g), respectively. CBF from ASL-MRI was significantly greater compared with CBF from H215O-PET (P < 0.001). However, these CBF values were significantly correlated with each other in the scatter plots (P < 0.05).
CONCLUSIONS: Noninvasive measurement of CBF using H215O-PET/MRI and IDIF with the cross-calibration method with a skull phantom experiment provided reasonable quantitative values. The IDIF method allowed reliable estimation of arterial radioactivity concentration, which is useful for clinical application. The ASL-MRI perfusion image from the simultaneous acquisition tended to overestimate CBF.