乳腺囊肿液分析与透射超声声速的相关性研究
本研究对37例乳腺大囊肿患者抽吸囊液,在体测量声速并检测电解质、蛋白、胆固醇、黏度和比重。结果发现黏度与胆固醇(r=0.71)、黏度与总蛋白×胆固醇(r=0.78)呈正相关,囊液细胞数与声速呈正相关(r=0.74)。作者提出乳腺大囊肿从液性到高细胞性的声速谱系,高聲速囊肿多为细胞数高、细胞团多的成熟大囊肿。
要点
- 乳腺大囊肿囊液细胞数与透射超声声速呈正相关(r=0.74)。
- 囊液黏度与胆固醇(r=0.71)及总蛋白×胆固醇(r=0.78)呈正相关。
- 高聲速囊肿可能为细胞数高、细胞团多的成熟大囊肿。
核心结论
- 作者提出乳腺大囊肿从液性到高细胞性的声速谱系,高聲速囊肿多为细胞数高、细胞团多的成熟大囊肿,与透射超声所见囊内微解剖相关。
研究设计与样本
- 病例收集研究,纳入37例乳腺超声成像患者的囊肿液样本,使用定量透射超声扫描仪在体测量声速,并检测囊液电解质、蛋白、胆固醇、黏度和比重。
关键影像表现/征象
- 透射超声可生成高分辨率三维声速图,用于识别乳腺组织类型并作为鉴别病变的生物标志物。
- 乳腺大囊肿的声速与囊内细胞团等细胞学特征相关。
诊断标准与关键数值
- 囊液细胞数与声速呈正相关(r=0.74)。
- 囊液黏度与胆固醇呈正相关(r=0.71)。
- 囊液黏度与总蛋白×胆固醇呈正相关(r=0.78)。
临床应用与阅片要点
- 作者认为声速测量可能有助于乳腺大囊肿的临床管理,但需进一步研究确认并评估其临床价值。
局限性
- 原文指出需进一步研究以确认本研究发现。
- 原文指出需进一步研究评估透射超声声速测量在乳腺成像中的临床价值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: The purpose of this work is to determine if the speed of sound value of a breast cyst can aid in the clinical management of breast masses. Breast macrocysts are defined as fluid-filled tissue masses >1 cm in diameter and are thought to be aberrations of normal development and involution, often associated with apocrine metaplasia. The benign natural history of breast cysts is well known, and it is important to obtain high specificity in breast imaging to avoid unnecessary biopsies in women who have benign diseases, particularly those with dense breast tissue. Transmission ultrasound is a tomographic imaging modality that generates high-resolution, 3D speed of sound maps that could be used to identify breast tissue types and act as a biomarker to differentiate lesions. We performed this study to investigate the microanatomy of macrocysts observed using transmission ultrasound, as well as assess the relationship of speed of sound to the physical and biochemical parameters of cyst fluids.
MATERIALS AND METHODS: Cyst fluid samples were obtained from 37 patients as part of a case-collection study for ultrasound imaging of the breast. The speed of sound of each sample was measured using a quantitative transmission ultrasound scanner in vivo. Electrolytes, protein, cholesterol, viscosity, and specific gravity were also measured (in the aspirated cyst fluid) to assess their relationship to the speed of sound values obtained during breast imaging.
RESULTS: We found positive correlations between viscosity and cholesterol (r = 0.71) and viscosity and total protein × cholesterol (r = 0.78). Additionally, we performed direct cell counts on cyst fluids and confirmed a positive correlation of number of cells with speed of sound (r = 0.74). The speed of sound of breast macrocysts, as observed using transmission ultrasound, correlated with the cytological features of intracystic cell clumps.
CONCLUSION: On the basis of our work with speed as a classifier, we propose a spectrum of breast macrocysts from fluid-filled to highly cellular. Our results suggest high-speed cysts are mature macrocysts with high cell counts and many cellular clumps that correlate with cyst microanatomy as seen by transmission ultrasound. Further studies are needed to confirm our findings and to assess the clinical value of speed of sound measurements in breast imaging using transmission ultrasound.