文献速递腹部技术与安全CT

常规肝脏CT中两种自适应统计迭代重建算法(ASiR与ASiR-V)与滤波反投影的图像质量比较

Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT.
PubMed 文献 · Br J Radiol · 2018-06-06

该研究纳入52例临床疑似肝转移患者,等分为ASiR组与ASiR-V组,扫描参数相同,分别以0(FBP)至100%以10%间隔的混合百分比重建,比较客观与主观图像质量。结果显示ASiR与ASiR-V较FBP均降低图像噪声并改善主观评分,ASiR-V可接受百分比更高,ASiR-V60%主观评分最高。

要点

  • ASiR与ASiR-V较FBP均显著降低图像噪声并改善主观图像质量评分。
  • ASiR主观评分在10%至80%区间显著改善,50%时达峰值,噪声降低32.2%。
  • ASiR-V主观评分在10%至90%区间显著改善,60%时达峰值,噪声降低51.5%,为最高评分。

核心结论

  • ASiR与ASiR-V较FBP均改善常规肝脏CT的客观与主观图像质量;ASiR-V提供进一步图像质量改善且可接受百分比高于ASiR,ASiR-V60%图像质量评分最高。

研究设计与样本

  • 经机构审查委员会批准的研究,纳入52例临床疑似肝转移患者,等分为ASiR组与ASiR-V组,两组扫描参数相同,分别以0(FBP)至100%以10%间隔的混合百分比重建,由两名有经验的放射科医师盲法独立评估图像质量,以FBP重建为参考标准。

诊断标准与关键数值

  • ASiR与ASiR-V百分比从10%增至100%时,图像噪声较FBP分别降低8.6–57.9%与8.9–81.6%;ASiR主观评分在10%至80%区间随百分比增加显著改善,峰值在50%且噪声降低32.2%;ASiR-V主观评分在10%至90%区间随百分比增加显著改善,峰值在60%且噪声降低51.5%;两名观察者间图像质量评估具有高度一致性。

临床应用与阅片要点

  • 常规肝脏CT重建时可参考ASiR与ASiR-V较FBP降低噪声并改善主观评分的表现。
  • ASiR-V可接受混合百分比高于ASiR,ASiR-V60%在本研究中主观图像质量评分最高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare image quality of two adaptive statistical iterative reconstruction (ASiR and ASiR-V) algorithms using objective and subjective metrics for routine liver CT, with the conventional filtered back projection (FBP) reconstructions as reference standards.

METHODS: This institutional review board-approved study included 52 patients with clinically suspected hepatic metastases. Patients were divided equally into ASiR and ASiR-V groups with same scan parameters. Images were reconstructed with ASiR and ASiR-V from 0 (FBP) to 100% blending percentages at 10% interval in its respective group. Mean and standard deviation of CT numbers for liver parenchyma were recorded. Two experienced radiologists reviewed all images for image quality blindly and independently. Data were statistically analyzed.

RESULTS: There was no difference in CT dose index between ASiR and ASiR-V groups. As the percentage of ASiR and ASiR-V increased from 10 to 100% , image noise reduced by 8.6 -57.9% and 8.9-81.6%, respectively, compared with FBP. There was substantial interobserver agreement in image quality assessment for ASiR and ASiR-V images. Compared with FBP reconstruction, subjective image quality scores of ASiR and ASiR-V improved significantly as percentage increased from 10 to 80% for ASiR (peaked at 50% with 32.2% noise reduction) and from 10 to 90% (peaked at 60% with 51.5% noise reduction) for ASiR-V.

CONCLUSION: Both ASiR and ASiR-V improved the objective and subjective image quality for routine liver CT compared with FBP. ASiR-V provided further image quality improvement with higher acceptable percentage than ASiR, and ASiR-V60% had the highest image quality score. Advances in knowledge: (1) Both ASiR and ASiR-V significantly reduce image noise compared with conventional FBP reconstruction. (2) ASiR-V with 60 blending percentage provides the highest image quality score in routine liver CT.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Chen LH, Jin C, Li JY, et al. Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT. Br J Radiol. 2018. PMID: 29848018. doi:10.1259/bjr.20170655

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。