文献速递介入外周血管神经DSA

介入放射学中高辐射剂量病例的监测与随访

Monitoring and Follow-Up of High Radiation Dose Cases in Interventional Radiology.
PubMed 文献 · Acad Radiol · 2018-06-19

本研究在单中心安装网络辐射剂量监测软件,对参考点空气比释动能(Ka,r)>5000 mGy的介入放射学(IR)和神经介入放射学(NIR)操作进行标记、皮肤剂量估算与临床随访。10例(2.7%)IR和52例(5.1%)NIR超过估计皮肤剂量5000 mGy,NIR中5/49(10.2%)报告暂时性皮肤/毛发损伤,IR无组织反应报告。结论认为软件监测与记录可成功实施,辐射诱导组织反应相对不常见。

要点

  • 参考点空气比释动能(Ka,r)>5000 mGy可作为介入操作高剂量标记阈值,用于触发皮肤剂量估算与临床随访。
  • 368例IR中10例(2.7%)、1025例NIR中52例(5.1%)超过估计皮肤剂量5000 mGy,NIR高剂量比例更高。
  • 49例接受介入治疗的NIR患者中5例(10.2%)报告暂时性皮肤/毛发损伤,IR高剂量病例无组织反应报告。

核心结论

  • 介入操作中辐射剂量的软件监测与记录可以成功实施。
  • 辐射诱导的组织反应相对不常见。

研究设计与样本

  • 本研究为在本机构安装网络辐射剂量监测软件并实施标记流程后,对医院一年内全部IR和NIR操作进行的评估。
  • 排除2189例Ka,r<2000 mGy的IR操作后,进一步分析368例IR和1025例NIR操作。
  • 对Ka,r>5000 mGy的操作进行皮肤剂量估算、生成正式报告并由医生启动临床随访,同时回顾性查阅病历评估患者报告的组织反应。

诊断标准与关键数值

  • 以参考点空气比释动能(Ka,r)>5000 mGy作为标记高剂量操作并触发后续评估的阈值。
  • 对标记病例估算入口皮肤剂量,并以估计皮肤剂量5000 mGy为高剂量判定标准。
  • 368例IR中10例(2.7%)超过估计皮肤剂量5000 mGy,1025例NIR中52例(5.1%)超过该阈值。

临床应用与阅片要点

  • 对Ka,r>5000 mGy的IR和NIR操作应进行皮肤剂量估算并生成正式报告,以便医生启动临床随访。
  • 高剂量IR病例均为腹部/盆腔创伤血管造影(伴或不伴栓塞),未报告组织反应。
  • NIR高剂量病例中49例为介入操作、3例为诊断性血管造影,5例(10.2%)报告暂时性皮肤/毛发损伤。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To assess the implementation of radiation dose monitoring software, create a process for clinical follow-up and documentation of high-dose cases, and quantify the number of patient reported radiation-induced tissue reactions in fluoroscopically guided interventional radiology (IR) and neuro-interventional radiology (NIR) procedures.

MATERIALS AND METHODS: Web-based radiation dose monitoring software was installed at our institution and a process to flag all procedures with reference point air kerma (Ka,r) > 5000 mGy was implemented. The entrance skin dose was estimated and formal reports generated, allowing for physician-initiated clinical follow-up. To evaluate our process, we reviewed all IR and NIR procedures performed at our hospital over a 1-year period. For all procedures with Ka,r > 5000 mGy, retrospective medical chart review was performed to evaluate for patient reported tissue reactions.

RESULTS: Three thousand five hundred eighty-two procedures were performed over the 1-year period. The software successfully transferred dose data on 3363 (93.9%) procedures. One thousand three hundred ninety-three (368 IR and 1025 NIR) procedures were further analyzed after excluding 2189 IR procedures with Ka,r < 2000 mGy. Ten of 368 (2.7%) IR and 52 of 1025 (5.1%) NIR procedures exceeded estimated skin doses of 5000 mGy. All 10 IR cases were abdominal/pelvic trauma angiograms with/without embolization; there were no reported tissue reactions. Of 52 NIR cases, 49 were interventions and 3 were diagnostic angiograms. Five of 49 (10.2%) NIR patients reported skin/hair injuries, all of which were temporary.

CONCLUSION: Software monitoring and documentation of radiation dose in interventional procedures can be successfully implemented. Radiation-induced tissue reactions are relatively uncommon.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Perry BC, Ingraham CR, Stewart BK, et al. Monitoring and Follow-Up of High Radiation Dose Cases in Interventional Radiology. Acad Radiol. 2018. PMID: 29934019. doi:10.1016/j.acra.2018.04.020

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。