比较四种商用心脏磁共振特征追踪算法评估左心室整体心肌应变的可重复性

Left ventricular global myocardial strain assessment comparing the reproducibility of four commercially available CMR-feature tracking algorithms.
PubMed 文献 · Eur Radiol · 2018-06-05

在45例受试者(15例对照、15例急性心肌梗死、15例扩张型心肌病患者)中比较四种商用软件包测量左心室整体径向、环向和纵向收缩期峰值应变(GRS、GCS、GLS)的厂商间、观察者间和观察者内可重复性,结果显示GRS和GLS在不同厂商间不可互换,观察者间和观察者内可重复性优秀,但部分厂商测量GRS的可重复性存在问题。

要点

  • GRS和GLS在不同厂商软件间不可互换,GCS未见厂商间显著差异。
  • 观察者间与观察者内ICC分别为0.828-0.991和0.902-0.997,可重复性优秀。
  • 部分厂商测量GRS的变异系数较大,GRS在厂商间的CV差异大。

核心结论

  • 在心脏功能范围宽的受试者队列中,GRS和GLS值在不同厂商间不可互换。
  • 尽管各厂商观察者间和观察者内可重复性优秀,部分厂商在可重复测量整体径向应变方面存在问题。

研究设计与样本

  • 研究纳入45例受试者,包括15例对照、15例急性心肌梗死和15例扩张型心肌病患者。
  • 比较四种商用软件包测量左心室整体径向、环向和纵向收缩期峰值应变(GRS、GCS、GLS)的厂商间、观察者间(两名读者)和观察者内可重复性。
  • 采用方差分析(ANOVA)评估厂商间差异,采用组内相关系数(ICC)和变异系数(CV)评估观察者间和阅读间差异。

关键影像表现/征象

  • 归一化舒张末期容积为91、77和119 ml/m2(中位数、Q1、Q3),射血分数为41 ± 14%,范围12-67%。
  • GLS值为13.9% ± 5.4%(3.9-23.8%),GCS值为12.2% ± 5.8%(1.0-25.1%),GRS值为32.0% ± 14.7(3.6-67.8%)。

诊断标准与关键数值

  • ANOVA显示厂商间GRS差异显著(p < 0.001),GLS差异显著(p = 0.018),GCS差异不显著(p = 0.379)。
  • 观察者间和观察者内可重复性均未发现显著偏倚。
  • 观察者间ICC范围为0.828-0.991,观察者内ICC范围为0.902-0.997。
  • 观察者间CV范围为4.0-28.8%,观察者内CV范围为2.8-27.7%。
  • GRS的CV在厂商间差异大,观察者间为5.2-28.8%,观察者内为2.8-27.7%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the reproducibility of cardiovascular magnetic resonance feature-tracking (CMR-FT) packages to assess global left ventricular (LV) myocardial strain.

METHODS: In 45 subjects (i.e. 15 controls, 15 acute myocardial infarction, 15 dilated cardiomyopathy patients), we determined inter-vendor, inter-observer (two readers) and intra-observer reproducibility of peak systolic global radial, circumferential and longitudinal strain (GRS, GCS and GLS, respectively) comparing four commercially available software packages. Differences between vendors were assessed with analysis of variance (ANOVA), between observers and readings with intraclass correlation coefficient (ICC) and coefficient of variation (CV).

RESULTS: The normalised end-diastolic volume was 91, 77 and 119 ml/m2 (median, Q1, Q3) and ejection fraction was 41 ± 14%, range 12-67%. Global longitudinal strain (GLS), global circumferential strain (GCS) and global radial strain (GRS) values were 13.9% ± 5.4% (3.9-23.8%), 12.2% ± 5.8% (1.0-25.1%) and 32.0% ± 14.7 (3.6-67.8%), respectively. ANOVA showed significant differences between vendors for GRS (p < 0.001) and GLS (p = 0.018), not for GCS (p = 0.379). No significant bias was found for both intra- and inter-observer variability. The ICC for inter- and intra-observer reproducibility ranged 0.828-0.991 and 0.902-0.997, respectively. The CV, however, ranged considerably, i.e. 4.0-28.8% and 2.8- 27.7% for inter- and intra-observer reproducibility, respectively. In particular, for GRS differences in CV values between vendors were large, i.e. 5.2-28.8% and 2.8-27.7%, for inter- and intra-observer reproducibility, respectively.

CONCLUSIONS: In a cohort of subjects with a wide range of cardiac performances, GRS and GLS values are not interchangeable between vendors. Moreover, although intra- and inter-observer reproducibility amongst vendors is excellent, some vendors encounter problems to reproducibly measure global radial strain.

KEY POINTS: • Different software packages are currently available for myocardial strain assessment using routinely acquired cine CMR images. • Global myocardial strain values are not interchangeable between vendors for global longitudinal and global radial strain. • Inter- and intra-observer reproducibility for global strain assessment is excellent. However, some vendors encounter problems to reproducibly measure global radial strain.