单源双能CT评估椎体压缩骨折骨髓水肿的前瞻性诊断准确性研究
前瞻性纳入70例50岁以上影像学疑似胸腰椎压缩骨折患者,行单源双能CT(80/135 kVp)与1.5T MRI(含T1WI和STIR)对比,以MRI为参考标准评估虚拟去钙图像对骨髓水肿的诊断准确性。结果显示双能CT敏感度72%、特异度70%,阅片者间一致性Fleiss kappa为0.40,作者认为其可检出骨髓水肿但需培训有素的阅片者。
要点
- 单源双能CT虚拟去钙图像诊断椎体骨髓水肿敏感度72%、特异度70%,阅片者间一致性Fleiss kappa为0.40。
- MRI的Fleiss kappa为0.58,高于双能CT;T1WI的SNR和CNR显著优于STIR、CT和虚拟去钙图像(p<.0001)。
- 双能CT对既往手术干预椎体的骨髓水肿检出敏感度有限,阅片需有经验的医师。
核心结论
- 虚拟去钙图像可检出骨髓水肿且可在单源系统上采集,图像质量可接受,但敏感度72%、特异度70%、阅片者间一致性Fleiss kappa 0.40,需要培训有素的阅片者进行图像判读。
研究设计与样本
- 前瞻性诊断准确性研究,纳入70例(23男、47女,平均年龄70.7岁,SD 9.8)50岁以上影像学疑似胸腰椎压缩骨折患者,共分析192个椎体压缩骨折。
关键影像表现/征象
- 所有患者接受320排探测器CT单源双能序贯采集80和135 kVp数据集,并接受1.5T MRI包括T1加权和短tau反转恢复(STIR)序列。
- 虚拟去钙图像采用三物质分解算法重建,CT上存在高度丢失的椎体在MRI和双能CT上均进行骨髓水肿评分。
- T1加权图像的信号噪声比和对比噪声比显著优于STIR、CT和虚拟去钙图像(p<.0001),STIR与虚拟去钙图像之间无差异。
诊断标准与关键数值
- 以MRI为标准参考,双能CT检出骨髓水肿的敏感度为72%、特异度为70%(阅片者依赖)。
- 双能CT的Fleiss kappa为0.40,MRI的Fleiss kappa为0.58。
- 双能CT对既往手术干预椎体的骨髓水肿检出敏感度有限。
临床应用与阅片要点
- 双能CT虚拟去钙图像可用于检出椎体压缩骨折患者的骨髓水肿,但阅片者需经培训且有经验。
- 判读时需注意双能CT与MRI相比阅片者间一致性较低且特异度有限,对既往手术干预椎体敏感度有限。
局限性
- 双能CT的阅片者间一致性低于MRI(Fleiss kappa 0.40 vs 0.58)。
- 双能CT特异度有限(70%)。
- 双能CT对既往手术干预椎体的骨髓水肿检出敏感度有限。
- 图像判读需要培训有素的阅片者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate the diagnostic accuracy of single-source dual-energy computed tomography (DECT) for the detection of bone marrow oedema (BME) in patients with vertebral compression fractures.
METHODS: Patients over 50 years of age with radiographically suspected vertebral compression fracture of the thoracic or lumbar spine were prospectively enrolled. All patients underwent DECT with sequential acquisition of 80 and 135 kVp datasets on a 320-row detector CT scanner and 1.5-Tesla magnetic resonance imaging (MRI) including T1-weighted and short-tau inversion recovery (STIR) sequences. Virtual non-calcium (VNCa) images were reconstructed using a three-material decomposition algorithm. Vertebrae with height loss in CT were scored for the presence of BME in both MRI and DECT and used to determine signal- and contrast-to-noise ratios (SNR and CNR). Contingency analysis using MRI as standard of reference and Fleiss's kappa were calculated. IRB approval was obtained.
RESULTS: In total 192 vertebral compression fractures in 70 patients (23 men, 47 women; mean age 70.7 years (SD 9.8)) were included in our analysis. DECT showed a reader-dependent sensitivity of 72% and specificity of 70% for BME. Fleiss's kappa was .40 for DECT and .58 for MRI. T1-weighted images had significantly better SNR and CNR compared to STIR, CT, and VNCa (p < .0001); however, there was no difference between STIR and VNCa.
CONCLUSIONS: VNCa images depict BME with adequate sensitivity and specificity and can be acquired on a single-source system. Image quality is adequate but trained readers are needed for image interpretation.
KEY POINTS: • Dual-energy CT in a single-source technique can help to detect bone marrow oedema in patients with vertebral compression fractures. • However, given the inferior inter-rater reliability and limited specificity compared to MRI, experienced readers are needed for image interpretation. • Dual-energy CT of the spine has limited sensitivity for the detection of bone marrow oedema in vertebra with previous surgical intervention.