新型光谱探测器CT虚拟单能重建减少牙种植体伪影
回顾性纳入38例牙种植体患者,行光谱探测器CT后重建常规图像与40–200 keV虚拟单能图像。高keV虚拟单能图像可降低金属伪影并改善软腭和面颊评估,观察者间一致性良好(ICC=0.77)。
要点
- 高keV虚拟单能图像(≥100 keV)可显著减少牙种植体高密度金属伪影。
- 虚拟单能200 keV较常规图像降低脂肪区图像噪声并改善软腭与面颊显示。
- 观察者间一致性好(ICC=0.77)。
核心结论
- 光谱探测器CT虚拟单能图像可减少牙种植体金属伪影并改善周围软组织诊断评估。
研究设计与样本
- 回顾性研究纳入38例连续牙种植体患者,使用光谱探测器CT(IQon,Philips,Best,荷兰)扫描,重建常规图像及40–200 keV(10 keV间隔)虚拟单能图像。
关键影像表现/征象
- 低密度伪影在常规图像为-218.7 HU、虚拟单能200 keV为-174.4 HU(p=0.1)。
- 高密度伪影在常规图像为309.8 HU、虚拟单能200 keV为119.2 HU(p≤0.05)。
- 脂肪区图像噪声在常规图像为23.9、虚拟单能200 keV为16.4(p≤0.05)。
诊断标准与关键数值
- 高密度伪影评分在虚拟单能≥100 keV显著降低(常规图像与虚拟单能200 keV分别为2(1-3)与3(1-5),p≤0.05)。
- 软腭评估评分常规图像与虚拟单能200 keV分别为2(1-4)与3(1-5),p≤0.05。
- 面颊评估评分常规图像与虚拟单能200 keV分别为2(1-5)与3(1-5),p≤0.05。
- 观察者间一致性ICC=0.77。
临床应用与阅片要点
- 高keV虚拟单能图像(≥100 keV)可减少牙种植体高密度伪影,改善软腭及面颊软组织显示。
- 200 keV虚拟单能图像可降低脂肪区图像噪声并改善伪影区域的主观评估。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Image quality in head and neck imaging is often severely hampered by artifacts arising from dental implants. This study evaluates metal artifact (MA) reduction using virtual monoenergetic images (VMI) compared to conventional CT images (CI) from spectral-detector computed tomography (SDCT).
METHODS: 38 consecutive patients with dental implants were included in this retrospective study. All examinations were performed using a SDCT (IQon, Philips, Best, The Netherlands). Images were reconstructed as conventional images (CI) and as VMI in a range of 40-200 keV (10 keV increment). Quantitative image analysis was performed ROI-based by measurement of attenuation (HU) and standard deviation in most pronounced hypo- and hyperdense artifact, fat and soft tissue with presence of artifacts. Qualitatively, extent of artifact reduction, assessment of soft palate and cheeks were rated on 5-point Likert-scales by two radiologists. Statistical data evaluation included ANOVA and Wilcoxon-test with correction for multiple comparisons; interrater-agreement was determined by intraclass-correlation coefficient (ICC).
RESULTS: The hypo- and hyperattenuating artifacts showed an increase and decrease of HU-values in VMIhigh (CI/VMI200 keV: -218.7/-174.4 HU, p = 0.1; and 309.8/119.2, p ≤ 0.05, respectively). Artifacts in the fat, as depicted by image noise did also decrease in VMIhigh (CI/VMI200 keV: 23.9/16.4, p ≤ 0.05). Qualitatively, hyperdense artifacts were decreased significantly in VMI ≥100 keV (e.g. CI/VMI200 keV: 2(1-3)/3(1-5), p ≤ 0.05). Artifact reduction resulted in improved assessment of the soft palate and cheeks (e.g. CI/VMI200 keV: 2(1-4)/3(1-5) and 2(1-5)/3(1-5), p ≤ 0.05). Overall interrater agreement was good (ICC = 0.77).
CONCLUSIONS: Virtual monoenergetic images from SDCT reduce metal artifacts from dental implants and improve diagnostic assessment of surrounding soft tissue.