全脑灌注CT中使用眼眶屏蔽:有用防护还是障碍?
回顾性分析20例患者28次使用眼眶屏蔽的灌注CT检查,评估图像质量与伪影,并比较iMAR算法的效果。所有源图像均出现屏蔽相关伪影,85.7%达后颅窝;92.8%灌注图有伪影,多为错误色码。iMAR后93%灌注图伪影范围增加,图像质量无改善。结论为全脑灌注CT不推荐使用眼眶屏蔽。
要点
- 眼眶屏蔽在源图像和灌注图中均产生伪影,可累及后颅窝。
- 屏蔽层面区域因伪影而无法诊断。
- iMAR算法未能改善图像质量,反使93%灌注图伪影范围增加。
核心结论
- 全脑灌注CT中使用眼眶屏蔽会产生显著伪影,使屏蔽层面区域无法诊断,不推荐使用;建议采用其他降低辐射剂量的方法。
研究设计与样本
- 回顾性研究,纳入20例患者共28次使用眼眶屏蔽的灌注CT检查,分析源图像和灌注图,并与iMAR处理后的图像比较。
关键影像表现/征象
- 所有源图像均显示与眼眶屏蔽相关的伪影,85.7%的伪影达后颅窝。
- 92.8%的灌注图出现伪影,最常见为错误色码;有伪影的灌注图中92.3%达后颅窝。
- 源图像伪影类型包括噪声、条纹、线束硬化或上述类型的组合;灌注图伪影类型包括噪声、图像变形、色码缺失区域或错误色码。
临床应用与阅片要点
- 全脑灌注CT不推荐使用眼眶屏蔽,因其伪影使屏蔽层面区域无法诊断。
- 建议采用其他降低辐射剂量的替代方法。
- iMAR算法未能改善图像质量,应用后93%灌注图伪影范围增加。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the image quality and artefacts resulting from in-plane orbit shielding during computed tomography (CT) perfusion and to assess the possibility of artefact reduction using the iterative metal artefact reduction (iMAR) algorithm.
MATERIALS AND METHODS: Twenty-eight perfusion CT examinations obtained from 20 patients using orbit shields were included in this study. Source images and perfusion maps were analysed retrospectively to detect the type and extent of artefacts. Original images were compared with images processed using the iMAR algorithm. The extent of artefacts was categorised into three groups: orbital/frontal, middle fossa, and posterior fossa. Types of artefacts in source images were categorised as noise, streak, beam hardening, or a combination of those, and types of artefacts in perfusion maps were categorised as: noise, image distortion, areas with missing colour codes, or incorrect colour-coding.
RESULTS: All source images evaluated showed artefacts related to orbit shielding and 85.7% reached the posterior fossa. In 92.8% of scans, perfusion maps showed artefacts, mostly as incorrect colour codes. Of the perfusion maps with artefacts, 92.3% reached the posterior fossa. After application of the iMAR algorithm, an increase in the extent of artefacts was observed in 93% of perfusion maps, and no improvement of image quality was noted.
CONCLUSION: Orbit shields cause significant artefacts if used for in-plane shielding during whole-brain CT perfusion, and render areas at the level of the shield undiagnosable. Usage of an orbit shield during whole-brain CT perfusion is not recommended, so alternative methods for reducing the radiation dose are advisable.
引用
- Hakim A, Vulcu S, Dobrocky T, et al. Using an orbit shield during volume perfusion CT: is it useful protection or an obstacle? Clin Radiol. 2018. PMID: 29859633. doi:10.1016/j.crad.2018.05.003