评估自动千伏选择软件在儿童神经影像检查中降低晶状体剂量的作用
研究用1岁和5岁拟人模体及MOSFET探测器,在儿童脑部CT方案下比较自动千伏选择软件(120/110/100 kVp)与常规120 kVp的晶状体剂量和图像质量。1岁模体剂量降低9.8%、17.4%、19.6%,5岁模体降低1.2%、8.6%、17.3%,所有设置CNR均在对照11%以内。作者认为该软件可有效降低儿童晶状体剂量且图像质量可接受。
要点
- 自动千伏软件在1岁模体降低晶状体剂量达19.6%,5岁模体达17.3%。
- 所有自动千伏设置下CNR与对照相差不超过11%(1岁)和6%(5岁)。
- 剂量节省随管电压降低而增大,但5岁模体在120 kVp时仅降低1.2%。
核心结论
- 自动千伏选择软件可有效降低儿童CT扫描中眼晶状体的辐射剂量,且CNR保持在基线11%以内。
研究设计与样本
- 模体研究,使用两个拟人模体(1岁和5岁),以MOSFET探测器测量晶状体剂量,在临床儿童脑部方案下比较常规120 kVp与自动千伏软件(120、110、100 kVp)的剂量和CNR。
诊断标准与关键数值
- 常规脑扫描(无自动千伏)平均晶状体剂量:1岁模体0.92 ± 0.03 cGy,5岁模体0.81 ± 0.03 cGy。
- 自动千伏软件在120、110、100 kVp下,1岁模体剂量降低分别为9.8%、17.4%、19.6%,5岁模体分别为1.2%、8.6%、17.3%。
- 所有自动千伏扫描的CNR与对照扫描相差在1岁模体11%以内、5岁模体6%以内。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans.
MATERIALS AND METHODS: Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting.
RESULTS: The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom.
CONCLUSION: Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.