四维CT心内血流评估:与四维血流MRI的比较
该前瞻性研究纳入12名受试者,于2016年2月至12月间同期采集冠状动脉CTA和四维血流MRI数据,比较仅基于CT心脏解剖的血流模拟与四维血流MRI测量。四维血流CT获得的血流模式与四维血流MRI定性相似,峰值流率、每搏输出量、峰值收缩期积分动能及直接/残余血流成分定量结果高度相关,但早期和晚期充盈动能以及滞留/延迟成分无相关性。
要点
- 四维血流CT获得的血流模式与四维血流MRI定性相似(三名独立观察者分级,观察者内κ为0.83、0.79、0.70)。
- 峰值流率和每搏输出量在四维血流CT与四维血流MRI间高度相关(r=0.98和r=0.81,P<.05)。
- 峰值收缩期积分动能相关良好(r=0.95,P<.05),但早期和晚期充盈动能无相关性。
核心结论
- 四维血流CT产生与当前参考标准四维血流MRI定性及定量相似的心内血流模式。
研究设计与样本
- 前瞻性研究,纳入12名受试者(年龄范围36-74岁,平均57岁;男性7名,年龄范围36-74岁,平均57岁;女性5名,年龄范围52-73岁,平均64岁),于2016年2月至12月间采集冠状动脉CTA和四维血流MRI数据;仅基于CT心脏解剖进行血流模拟,并与四维血流MRI测量比较。
关键影像表现/征象
- 四维血流CT获得的血流模式与四维血流MRI测量定性相似,由三名独立观察者分级。
诊断标准与关键数值
- 观察者内变异采用Cohen κ评估,值为0.83、0.79和0.70;配对Wilcoxon秩和检验显示分级间无显著变化(P>.05)。
- 峰值流率和每搏输出量在四维血流MRI与四维血流CT间高度相关(r=0.98和r=0.81,P<.05)。
- 峰值收缩期积分动能相关良好(r=0.95,P<.05),而早期和晚期充盈动能水平无相关性。
- 血流成分分析显示直接成分和残余成分高度相关(r=0.93,P<.05和r=0.87,P<.05),而滞留成分和延迟成分无相关性。
临床应用与阅片要点
- 四维血流CT可用于评估心内血流模式,其定性和定量结果与四维血流MRI相似,但早期和晚期充盈动能及滞留/延迟血流成分与MRI无相关性。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To investigate four-dimensional (4D) flow CT for the assessment of intracardiac blood flow patterns as compared with 4D flow MRI. Materials and Methods This prospective study acquired coronary CT angiography and 4D flow MRI data between February and December 2016 in a cohort of 12 participants (age range, 36-74 years; mean age, 57 years; seven men [age range, 36-74 years; mean age, 57 years] and five women [age range, 52-73 years; mean age, 64 years]). Flow simulations based solely on CT-derived cardiac anatomy were assessed together with 4D flow MRI measurements. Flow patterns, flow rates, stroke volume, kinetic energy, and flow components were quantified for both techniques and were compared by using linear regression. Results Cardiac flow patterns obtained by using 4D flow CT were qualitatively similar to 4D flow MRI measurements, as graded by three independent observers. The Cohen κ score was used to assess intraobserver variability (0.83, 0.79, and 0.70) and a paired Wilcoxon rank-sum test showed no significant change (P > .05) between gradings. Peak flow rate and stroke volumes between 4D flow MRI measurements and 4D flow CT measurements had high correlation (r = 0.98 and r = 0.81, respectively; P < .05 for both). Integrated kinetic energy quantified at peak systole correlated well (r = 0.95, P < .05), while kinetic energy levels at early and late filling showed no correlation. Flow component analysis showed high correlation for the direct and residual components, respectively (r = 0.93, P < .05 and r = 0.87, P < .05), while the retained and delayed components showed no correlation. Conclusion Four-dimensional flow CT produced qualitatively and quantitatively similar intracardiac blood flow patterns compared with the current reference standard, four-dimensional flow MRI. © RSNA, 2018 Online supplemental material is available for this article.
引用
- Lantz J, Gupta V, Henriksson L, et al. Intracardiac Flow at 4D CT: Comparison with 4D Flow MRI. Radiology. 2018. PMID: 29944089. doi:10.1148/radiol.2018173017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。