文献速递腹部技术与安全CT

双层光谱CT碘定量的准确性:迭代重建、患者体型和管参数的影响

Accuracy of iodine quantification in dual-layer spectral CT: Influence of iterative reconstruction, patient habitus and tube parameters.
PubMed 文献 · Eur J Radiol · 2018-03-08

在腹部CT体模上评估双层光谱CT碘定量准确性受迭代重建、管电压和体型的影响。低CTDIvol(2.5 mGy)平均绝对百分比差异为34.72%,高CTDIvol(20 mGy)为5.89%;诊断剂量水平下可准确定量(MAPD 3.38%),迭代重建水平无显著影响,140 kVp下更准确。

要点

  • 诊断剂量水平下双层光谱CT碘定量准确,MAPD为3.38%。
  • 低剂量水平(2.5 mGy)碘定量准确性下降,MAPD为34.72%。
  • 迭代重建水平(i0-i7)对碘测量无显著影响。

核心结论

  • 双层光谱CT可准确定量低浓度碘,采用诊断辐射剂量时准确性更高。

研究设计与样本

  • 在双层光谱CT上扫描腹部CT体模,体模带不同延伸环和四个碘插入物(1、2、5、10 mg/ml),管电压120和140 kVp,CTDIvol为2.5、5、10和20 mGy,重建八个迭代重建水平(i0-i7)。

关键影像表现/征象

  • 低CTDIvol水平(2.5 mGy)MAPD为34.72%,高CTDIvol水平(20 mGy)MAPD为5.89%。
  • 诊断剂量水平下碘定量准确,MAPD为3.38%。
  • 管电压140 kVp下碘定量更准确。
  • 体模尺寸仅在低剂量水平有显著影响;诊断剂量水平下体模尺寸影响减小,所有体模尺寸MAPD<5%。

诊断标准与关键数值

  • 以平均绝对百分比差异(MAPD)评估碘定量准确性;低剂量2.5 mGy时MAPD为34.72%,高剂量20 mGy时MAPD为5.89%,诊断剂量水平MAPD为3.38%,所有体模尺寸在诊断剂量水平MAPD<5%。

临床应用与阅片要点

  • 采用诊断辐射剂量时双层光谱CT碘定量准确性更高。
  • 迭代重建水平不影响碘测量准确性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Evaluation of the influence of iterative reconstruction, tube settings and patient habitus on the accuracy of iodine quantification with dual-layer spectral CT (DL-CT).

MATERIAL AND METHODS: A CT abdomen phantom with different extension rings and four iodine inserts (1, 2, 5 and 10 mg/ml) was scanned on a DL-CT. The phantom was scanned with tube-voltages of 120 and 140 kVp and CTDIvol of 2.5, 5, 10 and 20 mGy. Reconstructions were performed for eight levels of iterative reconstruction (i0-i7). Diagnostic dose levels are classified depending on patient-size and radiation dose.

RESULTS: Measurements of iodine concentration showed accurate and reliable results. Taking all CTDIvol-levels into account, the mean absolute percentage difference (MAPD) showed less accuracy for low CTDIvol-levels (2.5 mGy: 34.72%) than for high CTDIvol-levels (20 mGy: 5.89%). At diagnostic dose levels, accurate quantification of iodine was possible (MAPD 3.38%). Level of iterative reconstruction did not significantly influence iodine measurements. Iodine quantification worked more accurately at a tube voltage of 140 kVp. Phantom size had a considerable effect only at low-dose-levels; at diagnostic dose levels the effect of phantom size decreased (MAPD <5% for all phantom sizes).

CONCLUSION: With DL-CT, even low iodine concentrations can be accurately quantified. Accuracies are higher when diagnostic radiation doses are employed.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Sauter AP, Kopp FK, Münzel D, et al. Accuracy of iodine quantification in dual-layer spectral CT: Influence of iterative reconstruction, patient habitus and tube parameters. Eur J Radiol. 2018. PMID: 29685549. doi:10.1016/j.ejrad.2018.03.009

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。