颈动脉纵切面角度无关主应变成像的体模与在体评估
研究提出利用复合平面波成像计算主应变弹性图,可在颈动脉纵切面可靠显示与探头角度无关的应变分布;体模中非平行角度下主应变均值与0°平行构型相差在2.5%以内,在体20名健康志愿者中弓形动脉的轴向应变显著低于主应变(p <0.01)。
要点
- 复合平面波成像的主应变弹性图可显示颈动脉纵切面角度无关的应变分布。
- 体模中±30°、±20°、±10°下主应变均值与0°相差在2.5%以内,轴向应变随角度增大而减小。
- 在体弓形动脉轴向应变显著低于主应变(p <0.01),与体模结果一致。
核心结论
- 体模与在体研究结果显示,复合平面波成像计算的主应变弹性图可可靠显示颈动脉纵切面角度无关的血管应变。
研究设计与样本
- 体模研究采用均匀冷冻凝胶血管体模,在探头角度0°、±10°、±20°、±30°下评估纵向平面径向变形可视化的可行性。
- 在体研究纳入20名50-60岁健康志愿者,使用商用超声扫描仪(Sonix RP,Ultrasonix Medical Corp.,Richmond,BC,Canada),发射频率5 MHz、采样频率40 MHz。
关键影像表现/征象
- 体模中管腔与探头平行的直血管,主应变与轴向应变相似。
- 体模非平行构型(±30°、±20°、±10°)下主应变均值幅度与平行构型(0°)相差在2.5%以内,相对不受角度变化影响;轴向应变幅度随角度增大而减小。
- 在体研究中,动脉较直者主应变与轴向应变弹性图相似;弓形动脉轴向应变显著低于相应主血管应变(p <0.01),与体模结果一致。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Non-invasive vascular elastography can evaluate the stiffness of the carotid artery by visualizing the vascular strain distribution. Axial strain estimates of the longitudinal cross section of the carotid artery are sensitive to the angle between the artery and the transducer. Anatomical variations in branching and arching of the carotid artery can affect the assessment of arterial stiffness. In this study, we hypothesized that principal strain elastograms computed using compounded plane wave imaging can reliably visualize the strain distribution in the carotid artery, independent of the transducer angle. We corroborated this hypothesis by conducting phantom and in vivo studies using a commercial ultrasound scanner (Sonix RP, Ultrasonix Medical Corp., Richmond, BC, Canada). The phantom studies were conducted using a homogeneous cryogel vessel phantom. The goal of the phantom study was to assess the feasibility of visualizing the radial deformation in the longitudinal plane of the vessel phantom, independent of the transducer angle (±30°, ±20°, ±10° and 0°). The in vivo studies were conducted on 20 healthy human volunteers in the age group 50-60 y. All echo imaging was performed at a transmit frequency of 5 MHz and sampling frequency of 40 MHz. The elastograms obtained from the phantom study revealed that for straight vessels, which had their lumen parallel to the transducer, principal strains were similar to axial strains. At non-parallel configurations (angles ±30°, ±20° and ±10°), the magnitudes of the mean principal strains were within 2.5% of the parallel configuration (0° angle) estimates and, thus, were observed to be relatively unaffected by change in angle. However, in comparison, the magnitude of the axial strain decreased with increase in angle because of coordinate dependency. Further, the pilot in vivo study indicated that the principal and axial strain elastograms were similar for subjects with relatively straight arteries. However, for arteries with arched geometry, axial strains were significantly lower (p <0.01) than the corresponding principal vascular strains, which was consistent with the results obtained from the phantom study. In conclusion, the results of the phantom and in vivo studies revealed that principal strain elastograms computed using CPW imaging could reliably visualize angle-independent vascular strains in the longitudinal plane of the carotid artery.
引用
- Nayak R, Schifitto G, Doyley MM, et al. Visualizing Angle-Independent Principal Strains in the Longitudinal View of the Carotid Artery: Phantom and In Vivo Evaluation. Ultrasound Med Biol. 2018. PMID: 29685590. doi:10.1016/j.ultrasmedbio.2018.03.012