双能CT单能重建图像在膝关节和髋关节假体金属伪影减少中的应用与优势
本研究评估双能CT(管电压100和140 kVp)减少31例患者共22个膝关节假体和10个髋关节假体金属伪影的图像质量,手动选择最佳单能级(64至190 keV)并与B60-140及Fast DE重建比较。两名放射科医师双盲定性评估的观察者间一致性接近100%,Fast DE重建整体图像质量最佳(37.8%评为优),其次为B60-140(20.5%评为优)和最佳单能级(10.2%评为优),64、69和88 keV单能图像评分最差。结论认为特定后处理的双能CT可减少金属伪影并显著提升图像质量和诊断价值。
要点
- Fast DE重建整体图像质量最佳,37.8%评为优,优于B60-140的20.5%和最佳单能级的10.2%。
- 64、69和88 keV单能图像评分最差,临床评估金属植入物时应避免优先选用。
- 双能CT特定后处理可减少金属伪影并提升图像质量和诊断价值。
核心结论
- 双能CT结合特定后处理可减少金属伪影,并显著提升评估金属植入物时的图像质量和诊断价值。
研究设计与样本
- 采用双能CT协议(管电压100和140 kVp)检查31例患者的22个膝关节假体和10个髋关节假体,单能重建 extrapolated 于64、69、88、105、110、120、140、170和190 keV并手动选择最佳能量,同时生成B60-140和Fast DE重建,由两名放射科医师双盲使用Likert量表定性评估,第三名放射科医师在每幅图像最明显条纹伪影处放置圆形ROI进行定量分析并以无伪影周围组织ROI为参考。
关键影像表现/征象
- 最佳单能级由操作者手动选择,具体能级未在摘要中报告。
- 64、69和88 keV单能图像评分最差。
- Fast DE重建整体图像质量最佳,37.8%评为优。
诊断标准与关键数值
- 定性评估使用Likert量表,两名放射科医师双盲评估,观察者间一致性接近100%。
- 整体图像质量评为优的比例:Fast DE为37.8%,B60-140为20.5%,最佳单能级为10.2%。
- 不同单能图像之间伪影数量有显著差异,但事后分析除170与64 keV比较呈接近统计学显著的趋势外未发现其他差异。
临床应用与阅片要点
- 评估膝关节和髋关节金属假体时,Fast DE重建整体图像质量最佳(37.8%评为优),可优先选用。
- 64、69和88 keV单能图像评分最差,阅片时应避免作为首选能级。
局限性
- 由于成对比较数量多,事后分析除170与64 keV比较呈接近统计学显著的趋势外未发现其他差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The study aimed to assess image quality when using dual-energy CT (DECT) to reduce metal artifacts in subjects with knee and hip prostheses.
METHODS: Twenty-two knee and 10 hip prostheses were examined in 31 patients using a DECT protocol (tube voltages 100 and 140 kVp). Monoenergetic reconstructions were extrapolated at 64, 69, 88, 105, 110, 120, 140, 170, and 190 kilo-electron volts (keV) and the optimal energy was manually selected. The B60-140 and Fast DE reconstructions were made by CT. The image quality and diagnostic value were subjectively and objectively determined. Double-blind qualitative assessment was performed by two radiologists using a Likert scale. For quantitative analysis, a circular region of interest (ROI) was placed by a third radiologist within the most evident streak artifacts on every image. Another ROI was placed in surrounding tissues without artifacts as a reference.
RESULTS: The inter-reader agreement for the qualitative assessment was nearly 100%. The best overall image quality (37.8% rated "excellent") was the Fast DE Siemens reconstruction, followed by B60-140 and Opt KeV (20.5 and 10.2% rated excellent). On the other hand, DECT images at 64, 69 and 88 keV had the worse scores. The number of artifacts was significantly different between monoenergetic images. Nevertheless, because of the high number of pairwise comparisons, no differences were found in the post hoc analysis except for a trend toward statistical significance when comparing the 170 and 64 keV doses.
CONCLUSIONS: DECT with specific post-processing may reduce metal artifacts and significantly enhance the image quality and diagnostic value when evaluating metallic implants.
引用
- Magarelli N, De Santis V, Marziali G, et al. Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses. Radiol Med. 2018. PMID: 29637389. doi:10.1007/s11547-018-0881-8