联合定性定量分析声辐射力脉冲成像评估乳腺病变的初步结果

Preliminary Results of Acoustic Radiation Force Impulse Imaging by Combined Qualitative and Quantitative Analyses for Evaluation of Breast Lesions.
PubMed 文献 · J Ultrasound Med · 2018-04-15

本研究纳入126个乳腺病灶,采用声辐射力脉冲成像的虚拟触诊组织成像(VTI)定性分型与剪切波速度定量分析,以病理为参考标准。VTI 4b型病灶均为恶性,评分系统以3~4分为恶性阈值,敏感度77.5%、特异度96.5%、准确率90.5%、AUC 0.933。

要点

  • VTI 4b型病灶均为恶性(20/20,50.0%),良性组93.0%病灶为1~3型。
  • 评分系统恶性阈值取3~4分时,敏感度77.5%、特异度96.5%、准确率90.5%、AUC 0.933。
  • 良性病灶VTI与B模式面积比、内部及边缘剪切波速度均显著低于恶性病灶(均P < .001)。

核心结论

  • ARFI技术可无创提供补充信息,可能成为鉴别乳腺良恶性病变的有效诊断工具,结果具前景。

研究设计与样本

  • 本研究为连续纳入100例女性患者、共126个乳腺病灶的初步研究,先行常规乳腺超声再行ARFI弹性成像,以病理分析为参考标准。
  • 采用Virtual Touch tissue imaging(VTI)与Virtual Touch tissue quantification(Siemens Medical Solutions,Mountain View,CA)分别定性、定量分析肿瘤的弹性与硬度。
  • 病理分析显示恶性组40个病灶、良性组86个病灶,并采用受试者工作特征曲线分析评估ARFI鉴别良恶性的诊断性能。

关键影像表现/征象

  • 良性与恶性乳腺病灶的VTI分型不同,良性组80个病灶(93.0%)为1、2或3型。
  • 所有4b型病灶(n = 20,50.0%)均为恶性。

诊断标准与关键数值

  • 用于评估ARFI诊断性能的评分系统,其恶性 cutoff 点估计在3~4分之间,对应敏感度77.5%、特异度96.5%、准确率90.5%、曲线下面积0.933。
  • 良性病灶的平均VTI与B模式面积比、内部剪切波速度及边缘剪切波速度均统计学显著低于恶性病灶(均P < .001)。

临床应用与阅片要点

  • 阅片时可结合VTI定性分型与剪切波速度定量测值综合判断乳腺病灶良恶性。
  • VTI 4b型病灶在本研究中均为恶性,良性病灶多表现为1~3型。
  • 评分系统恶性阈值位于3~4分之间,原文未给出更具体的单一阈值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study was to evaluate the application of a new elastographic technique, acoustic radiation force impulse (ARFI) imaging, and its diagnostic performance for characterizing breast lesions.

METHODS: One hundred consecutive female patients with 126 breast lesions were enrolled in our study. After routine breast ultrasound examinations, the patients underwent ARFI elasticity imaging. Virtual Touch tissue imaging (VTI) and Virtual Touch tissue quantification (Siemens Medical Solutions, Mountain View, CA) were used to qualitatively and quantitatively analyze the elasticity and hardness of tumors. A receiver operating characteristic curve analysis was performed to evaluate the diagnostic performance of ARFI for discrimination between benign and malignant breast lesions.

RESULTS: Pathologic analysis revealed 40 lesions in the malignant group and 86 lesions in the benign group. Different VTI patterns were observed in benign and malignant breast lesions. Eighty lesions (93.0%) of benign group had pattern 1, 2, or 3, whereas all pattern 4b lesions (n = 20 [50.0%]) were malignant. Regarding the quantitative analysis, the mean VTI-to-B-mode area ratio, internal shear wave velocity, and marginal shear wave velocity of benign lesions were statistically significantly lower than those of malignant lesions (all P < .001). The cutoff point for a scoring system constructed to evaluate the diagnostic performance of ARFI was estimated to be between 3 and 4 points for malignancy, with sensitivity of 77.5%, specificity of 96.5%, accuracy of 90.5%, and an area under the curve of 0.933.

CONCLUSIONS: The application of ARFI technology has shown promising results by noninvasively providing substantial complementary information and could potentially serve as an effective diagnostic tool for differentiation between benign and malignant breast lesions.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Wang L, Wan CF, Du J, et al. Preliminary Results of Acoustic Radiation Force Impulse Imaging by Combined Qualitative and Quantitative Analyses for Evaluation of Breast Lesions. J Ultrasound Med. 2018. PMID: 29656388. doi:10.1002/jum.14599

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。