超声检测乳腺手术标本中的微钙化
研究纳入20例因乳腺钙化行手术切除的女性,对手术标本行灰阶超声和MicroPure成像,4名盲法阅片者分析54个超声和54个MicroPure数字片段。阅片者用MicroPure看到的微钙化数量显著多于灰阶超声(14.0±12.0 vs 3.0±3.2,p<0.0001)。结论为在离体乳腺手术标本中MicroPure识别的乳腺微钙化多于灰阶超声。
要点
- MicroPure识别的微钙化数量显著多于灰阶超声(14.0±12.0 vs 3.0±3.2,p<0.0001)。
- 3名阅片者偏好MicroPure图像质量(p<0.009),1名阅片者偏好灰阶超声(p=0.003)。
- 3名阅片者用MicroPure见到的Cooper韧带伪影少于灰阶超声(p<0.0001),1名阅片者两者无显著差异(p=0.58)。
核心结论
- 在离体乳腺手术标本中,MicroPure识别的乳腺微钙化多于灰阶超声。
研究设计与样本
- 研究纳入20例计划行乳腺钙化区手术切除的女性,手术标本使用Aplio XG扫描仪(Canon)行灰阶超声和MicroPure检查,4名独立盲法阅片者分析54个超声和54个MicroPure数字片段。
关键影像表现/征象
- 所有阅片者用MicroPure看到的微钙化数量显著多于灰阶超声,14.0±12.0 vs 3.0±3.2(p<0.0001)。
- 3名阅片者认为MicroPure图像质量优于灰阶超声(p<0.009),1名阅片者认为灰阶超声图像质量优于MicroPure(p=0.003)。
- 3名阅片者用MicroPure见到的Cooper韧带伪影少于灰阶超声(p<0.0001),1名阅片者两者无显著差异(p=0.58)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The objective was to evaluate a commercial image processing technique (MicroPure, Canon Medical Systems, Tustin, CA, USA) for detection of microcalcifications in breast surgical specimens. Twenty women scheduled for surgical excision of an area with breast calcifications were enrolled, their surgical specimens underwent grayscale ultrasound (US) and MicroPure examination using an Aplio XG scanner (Canon). Four independent and blinded readers analyzed 54 US and 54 MicroPure digital clips to determine the number of calcifications and scored image quality and artifacts on a 10-point scale. All readers saw significantly more microcalcifications with MicroPure than with US, 14.0 ± 12.0 versus 3.0 ± 3.2 (p <0.0001). Three readers preferred MicroPure image quality over that of US (p <0.009) and vice versa for one reader (p = 0.003). Three readers saw fewer Cooper's ligament artifacts with MicroPure than with US (p <0.0001); one reader saw no significance difference between them (p = 0.58). In conclusion MicroPure identified more breast microcalcifications than grayscale US in ex vivo surgical breast specimens.
引用
- Machado P, Eisenbrey JR, Stanczak M, et al. Ultrasound Detection of Microcalcifications in Surgical Breast Specimens. Ultrasound Med Biol. 2018. PMID: 29606493. doi:10.1016/j.ultrasmedbio.2018.02.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。