双层光谱探测器CT与双源双能CT在体外鉴别尿酸与非尿酸肾结石中的非劣效性评估
研究将57枚离体尿路结石置于圆柱体模水浴中,用双层光谱探测器CT(SDCT)和第二、三代双源双能CT(dsDECT)在匹配扫描参数下扫描,以X射线衍射分析为参考标准。SDCT与dsDECT均能将纯尿酸结石与非尿酸结石完全区分(AUC=1),SDCT光谱重建在体模模型中鉴别尿酸与非尿酸结石的性能与dsDECT相似。
要点
- SDCT与dsDECT鉴别尿酸与非尿酸结石的AUC均为1。
- 尿酸结石中位VMR为0.95-0.99,DER为1.00-1.02。
- 非尿酸结石中位VMR为1.4-4.1,DER为1.39-1.69。
核心结论
- 在体模模型中,SDCT光谱重建鉴别尿酸与非尿酸结石的性能与dsDECT相似。
研究设计与样本
- 离体研究,57枚提取的尿路结石置于圆柱体模水浴中,在匹配扫描参数下用SDCT(IQon,Philips Healthcare)及第二、三代dsDECT(Somatom Flash和Force,Siemens Healthcare)扫描,以X射线衍射分析为临床参考标准,用自定义3D生长区域分割工具逐像素分割结石进行统计分析。
诊断标准与关键数值
- SDCT记录每个结石的40/200、62/92和62/100中位虚拟单能比(VMR);dsDECT从Syngo Via肾脏结石应用软件记录每个结石的双能比(DER);以AUC评估检出尿酸结石的诊断性能,SDCT与dsDECT的AUC均为1。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the non-inferiority of dual-layer spectral detector CT (SDCT) compared to dual-source dual-energy CT (dsDECT) in discriminating uric acid (UA) from non-UA stones.
METHODS: Fifty-seven extracted urinary calculi were placed in a cylindrical phantom in a water bath and scanned on a SDCT scanner (IQon, Philips Healthcare) and second- and third-generation dsDECT scanners (Somatom Flash and Force, Siemens Healthcare) under matched scan parameters. For SDCT data, conventional images and virtual monoenergetic reconstructions were created. A customized 3D growing region segmentation tool was used to segment each stone on a pixel-by-pixel basis for statistical analysis. Median virtual monoenergetic ratios (VMRs) of 40/200, 62/92, and 62/100 for each stone were recorded. For dsDECT data, dual-energy ratio (DER) for each stone was recorded from vendor-specific postprocessing software (Syngo Via) using the Kidney Stones Application. The clinical reference standard of X-ray diffraction analysis was used to assess non-inferiority. Area under the receiver-operating characteristic curve (AUC) was used to assess diagnostic performance of detecting UA stones.
RESULTS: Six pure UA, 47 pure calcium-based, 1 pure cystine, and 3 mixed struvite stones were scanned. All pure UA stones were correctly separated from non-UA stones using SDCT and dsDECT (AUC = 1). For UA stones, median VMR was 0.95-0.99 and DER 1.00-1.02. For non-UA stones, median VMR was 1.4-4.1 and DER 1.39-1.69.
CONCLUSION: SDCT spectral reconstructions demonstrate similar performance to those of dsDECT in discriminating UA from non-UA stones in a phantom model.
引用
- Ananthakrishnan L, Duan X, Xi Y, et al. Dual-layer spectral detector CT: non-inferiority assessment compared to dual-source dual-energy CT in discriminating uric acid from non-uric acid renal stones ex vivo. Abdom Radiol (NY). 2018. PMID: 29626256. doi:10.1007/s00261-018-1589-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。