PROPELLER-DWI在前列腺MRI中减少全髋关节置换金属伪影与变形的作用
该研究纳入21例有全髋关节置换史且临床怀疑或已知前列腺癌的男性,在1.5 T下比较单次激发EPI与PROPELLER(含/不含脂肪饱和)前列腺DWI的图像质量、伪影和变形。PROPELLER-DWI-FS较标准EP-DWI显著改善图像质量并减少变形(p<0.01),伪影未显著改善;末5例加做PROPELLER-DWI-NFS者伪影较EP-DWI显著减少(p<0.05),两种PROPELLER序列的定量变形均显著低于EP-DWI(p<0.01)。结论认为PROPELLER-DWI图像质量更好并减少伪影和变形。
要点
- 有全髋关节置换金属植入物时,PROPELLER-DWI-FS较标准EP-DWI显著改善图像质量并减少变形(p<0.01)。
- PROPELLER-DWI-NFS在末5例中较EP-DWI显著减少伪影(p<0.05),脂肪饱和序列伪影未见显著改善。
- 两种PROPELLER序列的定量变形均显著低于EP-DWI(p<0.01)。
核心结论
- 在有髋部金属植入物的患者中,PROPELLER-DWI较常规回波平面序列图像质量更好,并减少伪影和变形。
研究设计与样本
- 前瞻性影像学研究,纳入21例有全髋关节置换史且临床怀疑或已知前列腺癌的男性,平均年龄70.4岁、PSA 9.2 ng/ml;在1.5 T使用相控阵体线圈,采集单次激发EPI、PROPELLER-DWI-FS和PROPELLER-DWI-NFS,以T2WI为参考比较图像质量,并用T2W与DW图像体积生成定量变形图。
关键影像表现/征象
- 图像质量以5点Likert量表与T2WI比较,伪影按4点量表评估前列腺受累范围,变形按4点量表评估器官扭曲程度。
- 金属植入物位于双侧3例、左侧9例、右侧9例。
- 末5例额外采用PROPELLER-DWI-NFS成像。
诊断标准与关键数值
- 图像质量采用5点Likert量表与T2WI比较。
- 伪影采用4点量表,定义为前列腺受影响的范围;变形采用4点量表,定义为器官扭曲程度。
- PROPELLER-DWI-FS较标准EP-DWI图像质量改善(p<0.01)、变形减少(p<0.01),伪影未见显著改善。
- PROPELLER-DWI-NFS较EP-DWI伪影显著减少(p<0.05)。
- 定量变形:PROPELLER-DWI-FS较EP-DWI显著降低(p<0.01),PROPELLER-DWI-NFS较EP-DWI显著降低(p<0.01)。
临床应用与阅片要点
- 对既往全髋关节置换患者行前列腺DWI时,PROPELLER技术可改善图像质量并减少变形。
- 脂肪饱和PROPELLER-DWI可改善图像质量和变形,但对伪影的改善未达显著。
- 非脂肪饱和PROPELLER-DWI在末5例中显示伪影显著减少,样本量有限。
局限性
- PROPELLER-DWI-NFS仅在最后5例患者中额外采集。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare image quality, artefact, and distortion in standard echo-planar imaging (EPI) with periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) for prostate magnetic resonance imaging (MRI) diffusion-weighted imaging (DWI) in patients with previous total hip replacement (THR).
METHODS: 21 male subjects with a clinical suspicion for, or known prostate cancer and previous THR were scanned at 1.5 T using a phased-array body coil. DWI was obtained using single-shot EPI and PROPELLER techniques using fat saturation (PROPELLER-DWI-FS), and without (PROPELLER-DWI-NFS). Image quality (the overall impression of diagnostic quality) was compared to T2-weighted (T2WI) imaging using a 5-point Likert scale, with diffusion sequences additionally scored for artefact and distortion according to a 4-point scale, with artefact defined as the amount of prostate affected and distortion as the degree of warping of the organ. The T2W and DW image volumes were compared to produce quantitative distortion maps. A two-sample Wilcoxon test compared the qualitative scores, with inter-reader variability calculated using Cohen's kappa.
RESULTS: 21 patients were included in the study, with an average age of 70.4 years and PSA 9.2 ng/ml. Hip metalwork was present bilaterally in 3 patients, left-sided in 9, and right-sided in 9. PROPELLER-DWI-FS significantly improved image quality (p < 0.01) and reduced distortion (p < 0.01) when compared to standard EP-DWI. Artefact was not shown to be significantly improved. The last 5 patients in the study were additionally imaged with PROPELLER-DWI-NFS, which resulted in a significant reduction in artefact compared to EP-DWI (p < 0.05). Quantitative distortion was significantly lower compared to EP-DWI for both PROPELLER with fat saturation (p < 0.01) and without fat saturation (p < 0.01).
CONCLUSION: PROPELLER-DWI demonstrates better image quality and decreases both artefact and distortion compared to conventional echo planar sequences in patients with hip metalwork.
引用
- Czarniecki M, Caglic I, Grist JT, et al. Role of PROPELLER-DWI of the prostate in reducing distortion and artefact from total hip replacement metalwork. Eur J Radiol. 2018. PMID: 29685538. doi:10.1016/j.ejrad.2018.03.021
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。