本研究通过体模和患者腹部CT评估基于软件的肝表面结节性(LSN)评分的精确性。体模在单台CT上调整采集重建参数(34次)及4家厂商22台CT上扫描,患者组68例由12名读者比较软件与视觉评分。结果显示软件法重复性、可重复性、观察者间一致性和重测信度均优良。

要点

  • 软件法LSN评分在常规采集重建参数(kVp 100-140、mA 200-400、层厚1.25-5.0 mm等)下重复性优良(ICC 0.79-0.99)。
  • 22台不同CT扫描仪间可重复性优良:光滑段ICC 0.97(95%CI 0.95,0.99)、CV 7%;结节段ICC 0.94(95%CI 0.89,0.97)、CV 8%。
  • 观察者间一致性软件法优良(ICC 0.84;95%CI 0.79,0.88),优于视觉法(ICC 0.61;95%CI 0.51,0.69)。

核心结论

  • 基于软件的LSN评分是一种定量CT影像生物标志物,具有优良的重复性、可重复性、观察者间一致性和重测信度。

研究设计与样本

  • 构建含光滑与结节表面模拟肝及模拟内脏和皮下脂肪的拟人化CT体模。
  • 体模在单台CT扫描仪上调整图像采集和重建参数扫描多次(N=34),并在来自4家厂商、12个影像中心的22台不同CT扫描仪上扫描。
  • LSN评分通过基于软件的方法获得,采用组内相关系数(ICC)和变异系数评估重复性和可重复性。
  • 使用68例不同分期慢性肝病患者的腹部CT图像,由12名读者分别以软件法和视觉法评估LSN,采用ICC评估观察者间一致性和重测重复性。

诊断标准与关键数值

  • 常规图像采集和重建参数(kVp 100-140、mA 200-400及自动mA、层厚1.25-5.0 mm、视野35-50 cm、光滑或标准卷积核)下LSN评分重复性优良(ICC 0.79-0.99)。
  • 22台不同扫描仪CT图像LSN评分可重复性:光滑段ICC 0.97(95%CI 0.95,0.99)、CV 7%;结节段ICC 0.94(95%CI 0.89,0.97)、CV 8%。
  • 软件法LSN评分观察者间一致性优良(ICC 0.84;95%CI 0.79,0.88;CV 28%),视觉法良好(ICC 0.61;95%CI 0.51,0.69;CV 43%)。
  • 软件法LSN评分重测重复性优良(ICC 0.82;95%CI 0.79,0.84;CV 12%)。

临床应用与阅片要点

  • 软件法LSN评分可在常规采集重建参数下获得优良重复性,适用于不同CT扫描仪间的定量评估。
  • 软件法LSN评分观察者间一致性优于视觉法,阅片时采用软件定量评分可提高一致性。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate precision of a software-based liver surface nodularity (LSN) score derived from CT images.

METHODS: An anthropomorphic CT phantom was constructed with simulated liver containing smooth and nodular segments at the surface and simulated visceral and subcutaneous fat components. The phantom was scanned multiple times on a single CT scanner with adjustment of image acquisition and reconstruction parameters (N = 34) and on 22 different CT scanners from 4 manufacturers at 12 imaging centers. LSN scores were obtained using a software-based method. Repeatability and reproducibility were evaluated by intraclass correlation (ICC) and coefficient of variation. Using abdominal CT images from 68 patients with various stages of chronic liver disease, inter-observer agreement and test-retest repeatability among 12 readers assessing LSN by software- vs. visual-based scoring methods were evaluated by ICC.

RESULTS: There was excellent repeatability of LSN scores (ICC:0.79-0.99) using the CT phantom and routine image acquisition and reconstruction parameters (kVp 100-140, mA 200-400, and auto-mA, section thickness 1.25-5.0 mm, field of view 35-50 cm, and smooth or standard kernels). There was excellent reproducibility (smooth ICC: 0.97; 95% CI 0.95, 0.99; CV: 7%; nodular ICC: 0.94; 95% CI 0.89, 0.97; CV: 8%) for LSN scores derived from CT images from 22 different scanners. Inter-observer agreement for the software-based LSN scoring method was excellent (ICC: 0.84; 95% CI 0.79, 0.88; CV: 28%) vs. good for the visual-based method (ICC: 0.61; 95% CI 0.51, 0.69; CV: 43%). Test-retest repeatability for the software-based LSN scoring method was excellent (ICC: 0.82; 95% CI 0.79, 0.84; CV: 12%).

CONCLUSION: The software-based LSN score is a quantitative CT imaging biomarker with excellent repeatability, reproducibility, inter-observer agreement, and test-retest repeatability.