微泡增强超声对兔肝射频消融的影响

Effect of Microbubble-Enhanced Ultrasound on Radiofrequency Ablation of Rabbit Liver.
PubMed 文献 · Ultrasound Med Biol · 2018-04-21

研究在18只手术暴露的兔肝上评估微泡增强超声(MEUS)联合射频消融(RFA)的效果,另32只兔肝接受MEUS(n=14)或RFA(n=18)单独治疗作为对照。MEUS使超声造影峰值强度减半(44.7%-54.1%),联合组消融体积为RFA单独组的2.8倍,且无严重肝功能损害。

要点

  • MEUS预处理使肝血流灌注的超声造影峰值强度降低至44.7%-54.1%,即减半。
  • MEUS联合RFA的消融肝体积是RFA单独治疗的2.8倍。
  • MEUS治疗未造成严重肝功能损害。

核心结论

  • 兔肝RFA联合MEUS预处理可安全且大幅增强消融效果。

研究设计与样本

  • 动物实验:18只手术暴露兔肝行MEUS联合RFA,32只兔肝行MEUS(n=14)或RFA(n=18)单独治疗作为对照;MEUS使用831 kHz空气背衬聚焦换能器,以9 Hz频率发射400周期、4.3 MPa的脉冲,脂质微泡成核。

关键影像表现/征象

  • 超声造影证实MEUS治疗使肝血流灌注的造影峰值强度减半(44.7%-54.1%)。
  • MEUS联合RFA的消融肝体积为RFA单独治疗的2.8倍。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Microbubble-enhanced ultrasound (MEUS) can non-invasively disrupt and block liver blood perfusion. It may potentially overcome the heat sink effect during a thermal ablation and consequently enhance radiofrequency ablation (RFA) of the liver. We propose a new strategy combining RFA with MEUS. For ultrasound treatment, an 831-kHz air-backed focused transducer directed 400-cycle bursts at 4.3 MPa to the liver at a 9-Hz rate. The treatment was nucleated by a lipids microbubble forming MEUS. Eighteen surgically exposed rabbit livers were treated using MEUS combined with RFA; the other 32 livers were treated using MEUS (n = 14) or RFA (n = 18) alone and served as the controls. Contrast ultrasound imaging confirmed that MEUS treatment significantly reduced liver blood perfusion by cutting contrast peak intensities in half (44.7%-54.1%) without severe liver function damage. The ablated liver volume treated using MEUS combined with RFA was 2.8 times greater than that treated using RFA alone. In conclusion, RFA of the liver can be safely and greatly enhanced by combination with MEUS pre-treatment.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Chen Z, Zhao H, Qiao X, et al. Effect of Microbubble-Enhanced Ultrasound on Radiofrequency Ablation of Rabbit Liver. Ultrasound Med Biol. 2018. PMID: 29685588. doi:10.1016/j.ultrasmedbio.2018.03.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。