自动化与手动全身磁共振成像工作流程的性能比较
这项前瞻性研究纳入20例接受3T全身磁共振成像评估肌病的患者,比较自动化与手动工作流程。自动化流程显著缩短总检查时间(30.0±4.2 vs 41.5±3.4分钟,P<0.0001)和用户交互次数(10.2±4.4 vs 48.2±17.2,P<0.0001),图像质量与覆盖范围两组相当。
要点
- 自动化全身磁共振成像流程总检查时间30.0±4.2分钟,手动流程41.5±3.4分钟,自动化显著更短(P<0.0001)。
- 自动化流程用户交互次数10.2±4.4次,手动流程48.2±17.2次,自动化显著更低(P<0.0001)。
- 整体图像质量两组相似:TIRM序列自动化4.00±0.94 vs 手动3.45±1.19(P=0.264);ce-T1w序列自动化4.20±0.88 vs 手动4.55±.55(P=0.423)。
核心结论
- 自动化全身磁共振成像扫描仪工作流程较手动流程显著缩短检查时间和减少用户交互,图像质量与研究体积覆盖范围两组相当。
研究设计与样本
- 前瞻性研究,经当地伦理委员会批准,纳入20例因肌病评估接受3T全身磁共振成像的患者,10例采用原型自动化流程(7名女性,年龄52±13岁,体重69.9±13.3 kg,身高173±9.3 cm,BMI 23.2±3.0),10例采用手动扫描(6名女性,年龄35.9±12.4岁,体重72±21 kg,身高169.2±10.4 cm,BMI 24.9±5.6)。
诊断标准与关键数值
- 整体图像质量采用5分制评估(5分=优,1分=差),由2名阅片者对每个序列(冠状位T2加权涡轮反转恢复幅度序列[TIRM]和轴位增强T1加权梯度双回波序列)进行评估。
- 研究体积覆盖范围由2名阅片者评估。
- 阅片者间一致性采用组内相关系数评估:整体图像质量TIRM的ICC为0.95(95%CI 0.86-0.98),ce-T1w的ICC为0.88(95%CI 0.70-0.95)。
- MR技师对计划工作的接受度评分(1分=最高,10分=最低):自动化流程2.20±0.92 vs 手动流程4.80±2.39(P=0.005)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to evaluate the performance of an automated workflow for whole-body magnetic resonance imaging (WB-MRI), which reduces user interaction compared with the manual WB-MRI workflow.
MATERIALS AND METHODS: This prospective study was approved by the local ethics committee. Twenty patients underwent WB-MRI for myopathy evaluation on a 3 T MRI scanner. Ten patients (7 women; age, 52 ± 13 years; body weight, 69.9 ± 13.3 kg; height, 173 ± 9.3 cm; body mass index, 23.2 ± 3.0) were examined with a prototypical automated WB-MRI workflow, which automatically segments the whole body, and 10 patients (6 women; age, 35.9 ± 12.4 years; body weight, 72 ± 21 kg; height, 169.2 ± 10.4 cm; body mass index, 24.9 ± 5.6) with a manual scan. Overall image quality (IQ; 5-point scale: 5, excellent; 1, poor) and coverage of the study volume were assessed by 2 readers for each sequence (coronal T2-weighted turbo inversion recovery magnitude [TIRM] and axial contrast-enhanced T1-weighted [ce-T1w] gradient dual-echo sequence). Interreader agreement was evaluated with intraclass correlation coefficients. Examination time, number of user interactions, and MR technicians' acceptance rating (1, highest; 10, lowest) was compared between both groups.
RESULTS: Total examination time was significantly shorter for automated WB-MRI workflow versus manual WB-MRI workflow (30.0 ± 4.2 vs 41.5 ± 3.4 minutes, P < 0.0001) with significantly shorter planning time (2.5 ± 0.8 vs 14.0 ± 7.0 minutes, P < 0.0001). Planning took 8% of the total examination time with automated versus 34% with manual WB-MRI workflow (P < 0.0001). The number of user interactions with automated WB-MRI workflow was significantly lower compared with manual WB-MRI workflow (10.2 ± 4.4 vs 48.2 ± 17.2, P < 0.0001). Planning efforts were rated significantly lower by the MR technicians for the automated WB-MRI workflow than for the manual WB-MRI workflow (2.20 ± 0.92 vs 4.80 ± 2.39, respectively; P = 0.005). Overall IQ was similar between automated and manual WB-MRI workflow (TIRM: 4.00 ± 0.94 vs 3.45 ± 1.19, P = 0.264; ce-T1w: 4.20 ± 0.88 vs 4.55 ± .55, P = 0.423). Interreader agreement for overall IQ was excellent for TIRM and ce-T1w with an intraclass correlation coefficient of 0.95 (95% confidence interval, 0.86-0.98) and 0.88 (95% confidence interval, 0.70-0.95). Incomplete coverage of the thoracic compartment in the ce-T1w sequence occurred more often in the automated WB-MRI workflow (P = 0.008) for reader 2. No other significant differences in the study volume coverage were found.
CONCLUSIONS: In conclusion, the automated WB-MRI scanner workflow showed a significant reduction of the examination time and the user interaction compared with the manual WB-MRI workflow. Image quality and the coverage of the study volume were comparable in both groups.