常规120 kVp与光谱束成形150 kVp颞骨CT的辐射剂量与图像清晰度比较
研究比较150 kVp联合锡滤过光谱束成形与常规120 kVp颞骨CT的辐射剂量和图像清晰度,25例采用150 kVp-Sn方案、30例采用120 kVp方案。150 kVp-Sn方案有效剂量显著更低(0.26±0.26 mSv vs 0.92±0.10 mSv,p<0.001),23个解剖结构可分析性无显著差异,SNR略逊于120 kVp。
要点
- 150 kVp-Sn颞骨CT有效剂量0.26±0.26 mSv,显著低于120 kVp的0.92±0.10 mSv(p<0.001)。
- 23个颞骨解剖结构可分析性在150 kVp-Sn与120 kVp间无显著差异。
- 120 kVp图像SNR略优于150 kVp-Sn,但未影响解剖结构可分析性。
核心结论
- 150 kVp联合锡滤过光谱束成形行颞骨CT,在保持解剖结构清晰度的同时,较常规120 kVp显著降低辐射剂量,且低于既往文献低剂量方案。
研究设计与样本
- 前瞻性对照研究,150 kVp-Sn组25例(平均年龄46.8±21.2岁),120 kVp组30例(平均年龄54.5±17.8岁);150 kVp-Sn参数为200参考mAs自动管电流调制、64×0.6 mm准直、0.6 mm层厚、螺距0.8;120 kVp参数为180 mAs、128×0.6 mm准直、0.6 mm层厚、螺距0.8。
关键影像表现/征象
- 150 kVp-Sn组平均DLP和有效剂量为0.26±0.26 mSv,120 kVp组为0.92±0.10 mSv,150 kVp-Sn组显著更低(p<0.001)。
- 120 kVp图像SNR略优于150 kVp-Sn图像。
- 23个颞骨解剖结构的可分析性在150 kVp-Sn与120 kVp扫描间无显著差异。
诊断标准与关键数值
- 主观清晰度采用5点评分评估23个颞骨解剖结构。
- 客观清晰度采用4个不同ROI测量的SNR评估。
- 文献中最低有效剂量方案为0.31±0.12 mSv,作为螺旋多层颞骨CT低剂量方案参考。
临床应用与阅片要点
- 150 kVp联合锡滤过光谱束成形可在维持颞骨解剖结构可分析性的同时显著降低辐射剂量。
- 150 kVp-Sn方案有效剂量0.26±0.26 mSv,低于文献低剂量方案的0.31±0.12 mSv。
- 120 kVp图像SNR略优,但23个解剖结构可分析性无显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this article is to compare radiation doses and conspicuity of anatomic landmarks of the temporal bone between the CT technique using spectral beam shaping at 150 kVp with a dedicated tin filter (150 kVp-Sn) and the conventional protocol at 120 kVp.
METHODS: 25 patients (mean age, 46.8 ± 21.2 years) were examined using the 150-kVp Sn protocol (200 reference mAs using automated tube current modulation, 64 × 0.6 mm collimation, 0.6 mm slice thickness, pitch 0.8), whereas 30 patients (mean age, 54.5 ± 17.8 years) underwent the 120-kVp protocol (180 mAs, 128 × 0.6 mm collimation, 0.6 mm slice thickness, pitch 0.8). Radiation doses were compared between the two acquisition techniques, and dosimetric data from the literature were reviewed for comparison of radiation dose reduction. Subjective conspicuity of 23 anatomic landmarks of the temporal bone, expressed by 5-point rating scale and objective conspicuity by signal-to-noise ratio (SNR) which measured in 4 different regions of interest (ROI), were compared between 150-kVp Sn and 120-kVp acquisitions.
RESULTS: The mean dose-length-product (DLP) and effective dose were significantly lower for the 150-kVp Sn scans (0.26 ± 0.26 mSv) compared with the 120-kVp scans (0.92 ± 0.10 mSv, p < 0.001). The lowest effective dose from the literature-based protocols was 0.31 ± 0.12 mSv, which proposed as a low-dose protocol in the setting of spiral multislice temporal bone CT. SNR was slightly superior for 120-kVp images, however analyzability of the 23 anatomic structures did not differ significantly between 150-kVp Sn and 120-kVp scans.
CONCLUSION: Temporal bone CT performed at 150 kVp with an additional tin filter for spectral shaping markedly reduced radiation exposure when compared with conventional temporal bone CT at 120 kVp while maintaining anatomic conspicuity. The decreased radiation dose of the 150-kVp Sn was also lower in comparison to the previous literature-based low-dose temporal bone CT protocol.