基于模型的迭代重建(MBIR)对颅内动脉瘤治疗前后脑CT血管成像图像质量的影响

Impact of model-based iterative reconstruction (MBIR) on image quality in cerebral CT angiography before and after intracranial aneurysm treatment.
PubMed 文献 · Eur J Radiol · 2018-03-09

研究纳入107例患者(治疗前38例、治疗后69例),比较MBIR与ASIR重建的脑CT血管成像图像质量。客观评估显示MBIR降低噪声并提高SNR和CNR(均p<0.0001);主观评估显示未治疗组噪声、伪影、血管锐利度和整体质量均改善(p<0.0001),治疗组噪声和血管锐利度改善(p<0.0001),但金属伪影无变化。

要点

  • MBIR较ASIR降低图像噪声并提高SNR和CNR(均p<0.0001)。
  • 未治疗患者MBIR改善噪声、伪影、血管锐利度和整体质量(p<0.0001)。
  • 治疗后患者MBIR改善噪声和血管锐利度,但金属伪影无变化。

核心结论

  • MBIR显著改善颅内动脉瘤未治疗和治疗后患者的噪声、SNR、CNR和血管锐利度,但不减少治疗后金属装置产生的伪影。

研究设计与样本

  • 研究纳入107例患者(平均年龄58±14岁),其中颅内动脉瘤治疗前38例、治疗后69例。
  • 采用常规方案采集图像,分别以MBIR和ASIR重建。
  • 比较颈内动脉和大脑中动脉的图像噪声、SNR和CNR,使用Wilcoxon符号秩检验。
  • 由两名神经放射科医生用半定量评估量表主观评价噪声、伪影、血管锐利度和整体质量。

诊断标准与关键数值

  • 以ASIR为对照,MBIR在未治疗和治疗后患者中均降低噪声(p<0.0001)、提高SNR(p<0.0001)和CNR(p<0.0001)。
  • 未治疗患者主观评估:MBIR较ASIR改善噪声降低、伪影、血管锐利度和整体质量(p<0.0001)。
  • 治疗后患者主观评估:MBIR较ASIR改善噪声和血管锐利度(p<0.0001),伪影无变化。

临床应用与阅片要点

  • 阅片时可利用MBIR降低噪声、提高SNR和CNR的优势来评估颅内动脉。
  • 对颅内动脉瘤治疗后患者,MBIR虽改善噪声和血管锐利度,但不减少金属装置伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To subjectively and objectively assess the impact of model-based iterative reconstruction(MBIR) on image quality in cerebral computed tomography angiography compared to adaptive statistical iterative reconstruction (ASIR).

METHODS: 107 patients (mean age: 58 ± 14 years) were included prior to (n = 38) and after (n = 69) intracranial aneurysm treatment. Images were acquired using a routine protocol and reconstructed with MBIR and ASIR. Image noise, signal-to-noise (SNR) and contrast-to-noise (CNR) ratios in the internal carotid and middle cerebral arteries were compared between MBIR and ASIR using the Wilcoxon signed-rank test. Additionally, two neuroradiologists subjectively assessed noise, artefacts, vessel sharpness and overall quality using a semi-quantitative assessment scale.

RESULTS: Objective assessment revealed that MBIR reduced noise (p < 0.0001) and additionally improved SNR (p < 0.0001) and CNR (p < 0.0001) compared to ASIR in untreated and treated patients. Subjective assessment revealed that in untreated patients, MBIR improved noise reduction, artefacts, vessel sharpness and overall quality relative to ASIR (p < 0.0001). In the treated groups, noise and vessel sharpness were improved (p < 0.0001) with no change in artefacts on images reconstructed with MBIR compared to ASIR.

CONCLUSION: MBIR significantly improves noise, SNR, CNR and vessel sharpness in untreated and treated patients with intracranial aneurysms. MBIR does not reduce artefacts generated by metallic devices following intracranial aneurysm treatment.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Hajdu SD, Daniel RT, Meuli RA, et al. Impact of model-based iterative reconstruction (MBIR) on image quality in cerebral CT angiography before and after intracranial aneurysm treatment. Eur J Radiol. 2018. PMID: 29685523. doi:10.1016/j.ejrad.2018.03.011

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。