3T颅内血管壁MRI序列的比较
本研究在3T MR扫描仪上比较了6种颅内血管壁序列变体(扫描时间4:39至8:24分钟)与临床常用序列(6:42分钟)的图像质量和病灶检出。最快变体(4:39分钟)的血管壁SNR/CNR高于或相当于参考序列,且检出了参考序列上的所有血管壁病灶;作者结论为开发出扫描速度快30%且血管壁可见性良好的序列。
要点
- 最快变体(4:39分钟)较临床常用序列(6:42分钟)扫描时间缩短约30%。
- 最快变体血管壁SNR/CNR除一项外均高于6:42分钟变体(P < .008)。
- 最快变体(4:39分钟)检出了6:42分钟变体上的所有血管壁病灶。
核心结论
- 开发出扫描速度快30%的血管壁序列,具有高SNR/CNR,颅内血管壁可见性良好。
研究设计与样本
- 在3T MR扫描仪上测试6种序列变体(扫描时间4:39至8:24分钟),共扫描15名受试者:5名仅行平扫全部7种变体,另10名行平扫及增强的最快变体(4:39分钟)和临床常用变体(6:42分钟)。
关键影像表现/征象
- 定性评估显示最快变体(4:39分钟)与6:42分钟变体在平扫及增强时CNR受抑制最明显。
- 最快变体(4:39分钟)的SNR/CNR除一项外均高于6:42分钟变体(P < .008)。
- 最快变体(4:39分钟)检出了6:42分钟变体上识别的所有血管壁病灶。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Intracranial vessel wall MR imaging plays an increasing role in diagnosing intracranial vascular diseases. For a complete assessment, pre- and postcontrast sequences are required, and including other sequences, these result in a long scan duration. Ideally, the scan time of the vessel wall sequence should be reduced. The purpose of this study was to evaluate different intracranial vessel wall sequence variants to reduce scan duration, provided an acceptable image quality can be maintained.
MATERIALS AND METHODS: Starting from the vessel wall sequence that we use clinically (6:42 minutes), 6 scan variants were tested (scan duration ranging between 4:39 and 8:24 minutes), creating various trade-offs among spatial resolution, SNR, and contrast-to-noise ratio. In total, 15 subjects were scanned on a 3T MR imaging scanner: In 5 subjects, all 7 variants were performed precontrast-only, and in 10 other subjects, the fastest variant (4:39 minutes) and our clinically used variant (6:42 minutes) were performed pre- and postcontrast.
RESULTS: The fastest variant (4:39 minutes) had higher or comparable SNRs/contrast-to-noise ratios of the intracranial vessel walls compared with the reference sequence (6:42 minutes). Qualitative assessment showed that the contrast-to-noise ratio was most suppressed in the fastest variant of 4:39 minutes and the variant of 6:42 minutes pre- and postcontrast. SNRs/contrast-to-noise ratios of the fastest variant were all, except one, higher compared with the variant of 6:42 minutes (P < .008). Furthermore, the fastest variant (4:39 minutes) detected all vessel wall lesions identified on the 6:42-minute variant.
CONCLUSIONS: A 30% faster vessel wall sequence was developed with high SNRs/contrast-to-noise ratios that resulted in good visibility of the intracranial vessel wall.
引用
- Lindenholz A, Harteveld AA, Zwanenburg JJM, et al. Comparison of 3T Intracranial Vessel Wall MRI Sequences. AJNR Am J Neuroradiol. 2018. PMID: 29674412. doi:10.3174/ajnr.A5629