文献速递外周血管神经技术与安全US

健康成人颈动脉分叉高帧频向量血流成像与彩色多普勒成像的比较

High-Frame Rate Vector Flow Imaging of the Carotid Bifurcation in Healthy Adults: Comparison With Color Doppler Imaging.
PubMed 文献 · J Ultrasound Med · 2018-03-25

研究用平面波高帧频向量血流成像(VFI)与彩色多普勒成像对比评估60名健康志愿者60条颈动脉分叉的血流特征。VFI与彩色多普勒在96.7%病例中检测复杂血流一致,VFI显示复杂血流范围更大(47.7±28.5 mm² vs 29.2±19.9 mm²)且持续时间更长(380±218 ms vs 325±206 ms)。VFI在评估复杂血流模式方面比彩色多普勒更强大。

要点

  • VFI与彩色多普勒检测颈动脉复杂血流的一致率为96.7%,仅3.3%病例VFI单独发现短时小再循环区。
  • VFI显示的复杂血流范围大于彩色多普勒(47.7±28.5 mm² vs 29.2±19.9 mm²)。
  • VFI显示的复杂血流持续时间长于彩色多普勒(380±218 ms vs 325±206 ms)。

核心结论

  • VFI在检测血流紊乱方面与彩色多普勒成像同样有效,但在评估复杂血流模式方面更强大。

研究设计与样本

  • 研究纳入60名健康志愿者的60条颈动脉,使用配备基于平面波的高帧频向量血流成像(VFI)的商用系统进行定量和定性评估,并与彩色多普勒成像比较。

关键影像表现/征象

  • 颈内动脉(ICA)复杂血流与ΔICA窦部-颈总动脉(CCA)直径比的统计学显著差异相关。
  • VFI与彩色多普勒成像在96.7%的病例中检测复杂血流一致,在3.3%的病例中仅VFI识别出短持续时间的小再循环区。
  • VFI突出显示复杂血流的范围更大(均值±标准差47.7±28.5 mm²;中位数45.5 mm²),而彩色多普勒成像为(均值29.2±19.9 mm²;中位数29.5 mm²)。
  • VFI比彩色多普勒成像更好地描绘不同的复杂血流模式。
  • VFI显示血流紊乱持续时间更长(均值380±218毫秒;中位数352.5毫秒),而彩色多普勒成像为(均值325±206毫秒;中位数333毫秒)。

诊断标准与关键数值

  • ΔICA窦部-CCA直径比与复杂血流范围之间呈强相关(r=0.84)。
  • VFI与彩色多普勒成像测量的血流紊乱持续时间之间呈强相关(r=0.92)。

临床应用与阅片要点

  • VFI可用于评估颈动脉分叉的复杂血流模式,其显示的范围和持续时间均大于彩色多普勒成像。
  • VFI可检测到彩色多普勒成像未显示的短时小再循环区。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the carotid bifurcation in healthy adults using a commercial system equipped with high-frame rate vector flow imaging (VFI) based on the plane wave and to compare VFI with color Doppler imaging.

METHODS: Carotid bifurcation diameters and flow characteristics of 60 vessels in 60 healthy volunteers were evaluated quantitatively and qualitatively to assess complex flow patterns and their extension and duration.

RESULTS: Complex flow in the internal carotid artery (ICA) was associated with a statistically significant difference in the ΔICA sinus-to-common carotid artery (CCA) diameter ratio (the relative change in diameter between the CCA and ICA sinus.) Vector flow imaging and color Doppler imaging were in accordance when detecting complex flow in 96.7% of cases; in 3.3% of cases, only VFI identified small recirculation areas of short duration. Vector flow imaging highlighted a larger extension of the complex flow (mean ± SD, 47.7 ± 28.5 mm2 ; median, 45.5 mm2 ) compared with color Doppler imaging (mean, 29.2 ± 19.9 mm2 ; median, 29.5 mm2 ) and better depicted different complex flow patterns; a strong correlation (r = 0.84) was found between the ΔICA sinus-to-CCA diameter ratio and the complex flow extension. Vector flow imaging showed a longer duration of the flow disturbances (mean, 380 ± 218 milliseconds; median, 352.5 milliseconds) compared with color Doppler imaging (mean, 325 ± 206 milliseconds; median, 333 milliseconds), and there was a strong correlation (r = 0.92).

CONCLUSIONS: Vector flow imaging is as effective as color Doppler imaging in the detection of flow disturbances, but it is more powerful in the assessment of complex flow patterns.

来源:J Ultrasound Med,查看原文。版权归原作者所有。