低剂量设置对儿童透视引导介入操作中辐射暴露的影响

Impact of low dose settings on radiation exposure during pediatric fluoroscopic guided interventions.
PubMed 文献 · Eur J Radiol · 2018-01-09

该回顾性研究比较了降低探测器入射剂量前后儿童介入放射学操作中的辐射剂量,纳入0-18岁患者(降低前39例、降低后26例),涉及外周动静脉畸形栓塞、门静脉介入及经皮肝穿刺胆道引流(PTCD)。结果显示皮肤剂量与剂量面积乘积(DAP)均大幅下降,图像质量总体良好且两名阅片者间一致性好;作者结论为降低探测器入射剂量可显著降低辐射剂量负担。

要点

  • 降低探测器入射剂量后,动静脉畸形栓塞、门静脉介入及PTCD的皮肤剂量分别下降75.1%、80.5%和85.3%。
  • 对应的剂量面积乘积(DAP)分别下降82.5%、72.2%和79.8%。
  • 图像质量总体良好,低剂量设置实施前后差异不显著,两名阅片者间一致性好。

核心结论

  • 在儿童介入放射学操作中降低探测器入射剂量可显著降低辐射剂量负担。

研究设计与样本

  • 回顾性研究,比较降低探测器入射剂量前(n=39)与后(n=26)的0-18岁儿童介入放射学操作,操作类型包括外周动静脉畸形栓塞、门静脉介入及经皮肝穿刺胆道引流(PTCD),时间范围为2014年至2017年。

关键影像表现/征象

  • 降低探测器入射剂量后,动静脉畸形栓塞的皮肤剂量下降75.1%,门静脉介入下降80.5%,PTCD下降85.3%。
  • 降低探测器入射剂量后,动静脉畸形栓塞的DAP下降82.5%,门静脉介入下降72.2%,PTCD下降79.8%。
  • 图像质量总体良好,低剂量设置实施前后差异不显著,两名阅片者间一致性好。

临床应用与阅片要点

  • 低剂量设置实施后,操作室内手动切换至更高剂量水平的功能仍然可用,但并未比实施前更频繁地使用。
  • 作者结论为降低探测器入射剂量可显著降低儿童介入放射学操作的辐射剂量负担。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the effects of lowering the detector entrance exposure in children undergoing interventional radiology procedures.

MATERIALS AND METHODS: The study retrospectively investigated radiation dose levels in pediatric patients aged 0-18 years before (n = 39) and after (n = 26) lowering detector entrance dose, undergoing embolization of peripheral Arteriovenous malformations, Portal Vein Interventions or Percutaneous Transhepatic Cholangio Drainage (PTCD) between 2014 and 2017. Patient characteristics, fluoroscopy time, protocols used as well as resulting Skin Dose and Dose Area Product (DAP) were compared in each cohort. Image quality was assessed by two independent readers.

RESULTS: The two patient cohorts did not differ in terms of patient demographics. Similarly, fluoroscopy time did not differ before and after implementation of the low dose settings. An overall reduction of skin dose of 75.1% for AVM embolizations, 80.5% for Portal Vein Interventions and 85.3% for PTCD placement was observed. The DAP decrease was 82.5% for AVM embolizations, 72.2% for Portal Vein Interventions and 79.8% for PTCD placement. Image quality was generally considered to be good with an insignificant difference between pre and post implementation of the low dose approach and good agreement between the two readers. Manual inroom-switching to higher dose levels was possible, however this was not performed more frequently after implementation of the low dose settings.

CONCLUSION: Lowering the detector entrance dose in pediatric interventional radiology procedures results in a significant decrease of the radiation dose burden.

来源:Eur J Radiol,查看原文。版权归原作者所有。