文献速递技术与安全CT

CT质量保证:使用自动CT剂量监测系统实施更新的国家诊断参考水平

Quality assurance in CT: implementation of the updated national diagnostic reference levels using an automated CT dose monitoring system.
PubMed 文献 · Clin Radiol · 2018-03-20

研究评估德国联邦辐射防护办公室更新的CT诊断参考水平(DRL)通过自动CT剂量监测系统融入临床常规的情况,分析2016年成人27,103次CT检查的94,258个序列。实施后机构CTDIvol/DLP均值均低于DRL,骨盆、腰椎和面骨部分方案初始超阈,优化后剂量下降。自动监测系统有助于DRL实施过程的全面监测和辐射剂量优化。

要点

  • 自动CT剂量监测系统可全面监测DRL实施过程并辅助优化辐射剂量
  • 骨盆、腰椎和面骨部分方案初始CTDIvol/DLP超过国家DRL,优化后下降
  • 机构整体CTDIvol/DLP均值在实施和优化后均低于国家DRL

核心结论

  • 自动CT剂量监测系统不仅能全面监测DRL实施过程,还能帮助优化辐射剂量。

研究设计与样本

  • 研究分析2016年成人患者27,103次CT检查的94,258个CT序列,在将更新的国家DRL实施到临床常规前后对比CT辐射暴露,分析CTDIvol和DLP,并由三名放射科医生使用三点Likert量表评估主观图像质量。

关键影像表现/征象

  • 机构整体CTDIvol/DLP均值初始为DRL的65.2±32.9%/67.3±41.5%,优化后为59.4±32%/60.5±39.9%,均低于DRL。
  • 超过国家DRL的方案涉及骨盆(n=268;CTDIvol 107.7±65.7%/DLP 106.3±79.3%)、腰椎(n=91;160.8±74.7%/175.2±104.1%)和面骨(n=527;108±39%/152.7±75.7%)。
  • 优化后骨盆CTDIvol/DLP为87.9±73%/87.8±80.8%,腰椎为67.8±33.2%/74.5±50.6%,面骨为95.1±45.8%/133.3±74.6%。

诊断标准与关键数值

  • 以德国联邦辐射防护办公室更新的国家DRL为阈值,按解剖区域映射机构CT方案。
  • 主观图像质量采用三点Likert量表评估。

临床应用与阅片要点

  • 将机构CT协议按解剖区域映射到国家DRL,系统识别超阈方案并进行优化。
  • 利用自动CT剂量监测系统可全面追踪DRL实施前后的剂量变化。
  • 对骨盆、腰椎和面骨等超阈区域应重点优化扫描方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To evaluate the implementation of the updated computed tomography (CT) diagnostic reference levels (DRLs) from the German Federal Office for Radiation Protection into clinical routine using an automatic CT dose monitoring system.

METHODS AND MATERIALS: CT radiation exposure was analysed before and after implementing the updated national DRLs into routine clinical work in 2016. After the implementation process, institutional CT protocols were mapped to the anatomical regions for which DRLs were provided. Systematically, protocols that exceeded the thresholds were optimised and analysed in detail. The CT radiation output parameters analysed were volumetric CT dose index (CTDIvol) and dose-length product (DLP). Three radiologists evaluated subjective image quality using a three-point Likert scale.

RESULTS: The study included 94,258 CT series (from 27,103 CT examinations) in adult patients performed in 2016. When averaged over all body regions with available DRL, institutional CTDIvol/DLP values were always below the DRLs (65.2±32.9%/67.3±41.5% initially; 59.4±32%/60.5±39.9% after optimisation). Values exceeding the national DRLs were found for pelvis (n=268; CTDIvol 107.7±65.7%/DLP 106.3±79.3%), lumbar spine (n=91; 160.8±74.7%/175.2±104.1%), and facial bones (n=527; 108±39%/152.7±75.7%). After optimisation, CTDIvol and DLP were 87.9±73%/87.8±80.8% for the pelvis, 67.8±33.2%/74.5±50.6% for the lumbar spine and 95.1±45.8%/133.3±74.6% for the viscerocranium.

CONCLUSION: An automatic CT dose monitoring system enabled not only comprehensive monitoring of a DRL implementation process but can also help to optimise radiation exposure.

来源:Clin Radiol,查看原文。版权归原作者所有。