乳腺X线摄影图像质量与循证实践:数字乳腺摄影中乳房下角显示情况的分析
该回顾性定量研究随机选取BreastScreen NSW的2270对数字乳腺摄影图像,分析乳房下角(IMA)在内外斜位(MLO)上的显示率。结果显示IMA在单张图像中显示率为52.4%,在配对图像中为42.6%,低于全球筛查乳腺摄影认证标准所要求的50%-75%合规率。IMA显示率随年龄和压迫乳腺厚度增加而降低,随乳腺后乳头线(PNL)增加而升高。
要点
- 数字乳腺摄影MLO位中IMA的临床显示率(配对图像42.6%)低于认证标准要求的50%-75%。
- 年龄每增加1岁,IMA显示的比值降低2%;压迫乳腺厚度每增加1mm,IMA显示的比值降低1.6%。
- PNL每增加1mm,IMA显示的比值升高0.4%;年龄每增加1岁,PNL预测增加0.48mm。
核心结论
- IMA显示率(配对图像42.6%)临床可达到但远低于全球筛查乳腺摄影认证标准的合规要求(50%-75%),这些标准应与已报告的循证数据对齐。
研究设计与样本
- 回顾性定量研究,随机选取BreastScreen NSW的2270对数字乳腺摄影图像,通过直接测量和视觉分析收集数据,采用组内相关分析评估阅片者间和阅片者内信度。
诊断标准与关键数值
- IMA在单张图像中显示率为52.4%,在配对图像中显示率为42.6%。
- PNL与年龄呈线性关系(p值<0.001),年龄每增加1岁,PNL预测增加0.48mm。
- 年龄每增加1岁,IMA显示的比值降低2%(p值=0.001)。
- PNL每增加1mm,IMA显示的比值升高0.4%(p值=0.001)。
- 压迫乳腺厚度每增加1mm,IMA显示的比值降低1.6%(p值<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: The aim of this study is to determine the clinical rates of the demonstration of the inframammary angle (IMA) on the mediolateral oblique (MLO) view of the breast on digital mammograms and to compare the outcomes with current accreditation standards for compliance. Relationships between the IMA, age, the posterior nipple line (PNL) and compressed breast thickness will be identified and the study outcomes validated using appropriate analyses of inter-reader and inter-rater reliability and variability. Differences in left versus right data were also investigated.
METHOD: A quantitative retrospective study of 2270 randomly selected paired digital mammograms performed by BreastScreen NSW was undertaken. Data was collected by direct measurement and visual analysis. Intra-class correlation analyses were used to evaluate inter- and intra-rater reliability.
RESULTS: The IMA was demonstrated on 52.4% of individual and 42.6% of paired mammograms. A linear relationship was found between the posterior nipple line (PNL) and age (p-value <0.001). The PNL was predicted to increase by 0.48 mm for every one year increment in age. The odds of demonstrating the IMA reduced by 2% for every one year increase in age (p-value = 0.001); are 0.4% higher for every 1 mm increase in PNL (p-value = 0.001) and 1.6% lower for every 1 mm increase in compressed breast thickness, (p-value<0.001). There was high inter- and intra-rater reliability for the PNL while there was 100% agreement for the demonstration of the IMA.
CONCLUSION: Analysis of the demonstration of the IMA indicates clinically achievable rates (42.6%) well below that required for compliance (50%-75%) to known worldwide accreditation standards for screening mammography. These standards should be aligned to the reported evidence base. Visualisation of the IMA is impacted negatively by increasing age and compressed breast thickness but positively by breast size (PNL).