可变目标噪声水平的螺旋CT用于胸部、腹部和盆腔CT的剂量降低
研究纳入41例研究组患者行可变目标噪声水平(胸部SD=16、腹部/盆腔SD=13)的连续螺旋CT,并规划了两种未执行的方案作为对照;另纳入40例对照组采用两种螺旋扫描重叠于肝脏的方案。结果显示可变噪声方案较单次统一SD方案剂量降低17.4%,较两次分开扫描方案降低16.3%(均p<0.001),两组图像质量无差异(p>0.3)。
要点
- 胸部采用较高目标噪声(SD=16)、腹部/盆腔采用标准噪声(SD=13)的连续螺旋扫描可合并完成。
- 可变SD扫描较单次统一腹部SD方案剂量降低17.4%,较两次分开扫描方案降低16.3%(均p<0.001)。
- 可变SD扫描与两次分开扫描方案相比图像质量无差异(p>0.3)。
核心结论
- 胸部/腹部/盆腔连续螺旋扫描采用可变目标噪声水平,与单次仅用腹部剂量设置采集或两次分开采集胸部和腹部/盆腔相比,辐射剂量降低约17%,且图像质量无差异。
研究设计与样本
- 研究组41例患者行可变目标噪声水平(胸部SD=16、腹部/盆腔SD=13)的连续螺旋扫描(P1),另规划但未执行P2(全扫描范围单次螺旋扫描,仅SD=13)和P3(两次分开螺旋扫描,在肝脏区域重叠,SD设置同P1);对照组40例采用P3方案扫描。记录所有DLP,并对所有扫描进行定性和定量图像质量评估。
诊断标准与关键数值
- 研究组P1方案:胸部目标噪声SD=16,腹部/盆腔目标噪声SD=13。
- P1、P2、P3的DLP(均值/标准差)分别为859.5/392.9、1040.2/510.5、1027.4/469.4。
- P1较P2平均剂量降低17.4%(p<0.001),较P3平均剂量降低16.3%(p<0.001)。
- 两组患者图像质量无差异(p>0.3)。
局限性
- P2和P3方案为规划但未执行,未在实际患者中实施。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate a contiguous helical CT protocol with two different target noise levels in chest/abdomen/pelvis CT.
METHODS: 41 patients (study group) underwent a helical scan (P1) with two different target noise levels (SDs), SD = 16 for chest and SD = 13 for abdomen/pelvis. Two further protocols were planned but not executed: a single helical scan with only one SD (SD = 13) for the entire scan range (P2), and two separate helical scans overlapping over the liver and same SD settings as for P1 (P3). All DLPs were recorded. Image quality was assessed qualitatively and quantitatively on all scans. The control group consisted of 40 patients, was scanned with protocol P3 and analysed using the same metrics.
RESULTS: DLPs (mean/SD) for P1, P2 and P3 were 859.5/392.9, 1040.2/510.5 and 1027.4/469.4, respectively. P1 offered a mean dose reduction of 17.4% compared to P2, and 16.3% compared to P3 (both p < 0.001). There were no differences in image quality between both patient groups (p > 0.3).
CONCLUSION: Contiguous helical scanning of the chest/abdomen/pelvis with variable target noise levels results in approximately 17% dose reduction if compared to a single acquisition with only abdominal dose settings or two separate acquisitions of the chest and abdomen/pelvis.
KEY POINTS: • Low dose chest and standard abdomen CTs can be combined. • Variable SD CT scanning allows for radiation dose reduction. • Variable SD CT scanning maintains image quality.