文献速递外周血管介入技术与安全CTDSA

虚拟单源CT采用双源采集:在下肢流出道CT血管造影中个体内比较不同扫描协议的临床适用性

Virtual single source CT using dual source acquisition: Clinical applicability in run-off CT-angiography for intra-individual comparison of different scan protocols.
PubMed 文献 · Eur J Radiol · 2018-02-13

本研究为前瞻性研究,52例患者于2012年6月至2013年6月接受下肢流出道CT血管造影,利用双源CT的虚拟单源CT采集在一次扫描中重建三种辐射剂量水平(RDL40、RDL80、RDL120)的图像。RDL120图像噪声和CNR最佳,RDL40与RDL120在腹部至足部水平定性图像质量有显著差异;RDL40至RDL80在腹部和盆腔水平剂量加权CNR无显著差异,大腿至足部水平CNRD随剂量降低显著升高。结论认为VSS-CT可用于流出道CTA的个体内协议比较,并发现辐射剂量较制造商推荐协议降低约1/3的潜力。

要点

  • 虚拟单源CT可在一次双源采集下重建三个辐射剂量水平(RDL40、RDL80、RDL120)用于个体内比较。
  • RDL120图像噪声和CNR最佳,RDL80和RDL40显著降低(p < 0.001),与解剖水平无关。
  • 腹部和盆腔水平CNRD在各剂量间无显著差异,大腿至足部水平CNRD随剂量降低显著升高。

核心结论

  • VSS-CT的辐射剂量拆分可应用于流出道CTA,实现不同采集协议的个体内比较而不增加额外辐射暴露;本研究识别出较制造商推荐采集协议降低约1/3辐射剂量的潜力。

研究设计与样本

  • IRB批准的前瞻性研究,纳入52例于2012年6月至2013年6月接受流出道CTA的患者,使用第一代双源CT行VSS-CT采集,管电压120 kVp、准直2×32×0.6 mm、层厚1.0 mm,管A电流时间乘积80 mAs、管B 40 mAs,分别以管B、管A及双管联合原始数据重建RDL40、RDL80、RDL120三组图像。

诊断标准与关键数值

  • 平均DLP在RDL40、RDL80、RDL120分别为200±40、400±90、600±130 mGy·cm;图像噪声和CNR在RDL120最佳,RDL80和RDL40显著降低(p < 0.001);CNRD在腹部和盆腔水平各剂量间无显著差异,大腿至足部水平RDL120、RDL80、RDL40间显著升高;定性图像质量在RDL120与RDL40间从腹部至足部水平有显著差异,其他剂量水平间无差异。

临床应用与阅片要点

  • VSS-CT可在一次双源采集下重建不同辐射剂量水平的图像,便于个体内比较不同采集协议;腹部和盆腔水平CNRD在各剂量间无显著差异,而大腿至足部水平CNRD随剂量降低显著升高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Virtual single source computed tomography (VSS-CT) acquisition on a dual source CT (DSCT) has been demonstrated to allow for dose-neutral intra-individual comparison of three acquisition protocols at different radiation dose levels (RDL) within one acquisition in a phantom. The purpose of this study was twofold: first to evaluate the applicability of VSS-CT in patients and second to optimize the task-dependent trade-off between radiation dose and image quality of lower extremity CT angiography (run-off CTA).

MATERIAL AND METHODS: In this IRB-approved prospective study 52 patients underwent run-off CTA between 06/2012 and 06/2013. VSS-CT acquisition was conducted using a first generation DSCT applying equal X-ray tube settings (120 kVp), collimation (2 × 32 × 0.6 mm), and slice thickness (1.0 mm) but different effective tube current-time products (tube A: 80 mAs, tube B: 40 mAs). Three different image datasets representing three different radiation dose levels (RDL40, RDL80, RDL120) were reconstructed using a soft kernel from the raw data of tube B, tube A or both tubes combined. Dose length products (DLP) of each raw data set were documented. Quantitative image quality (IQ) was assessed for five anatomical levels using image noise and contrast-to-noise ratio (CNR). To investigate dose efficiency of each acquisition, the dose-weighted CNR (CNRD) was determined. Qualitative IQ was evaluated by two blinded readers in consensus using a 5-point Likert scale and compared with a Friedman- and posthoc Wilcoxon test.

RESULTS: Mean DLP was 200 ± 40, 400 ± 90 and 600 ± 130 mGy·cm for the RDL40, RDL80 and RDL120, respectively. Image noise and CNR were best for RDL120 and decreased significantly for RDL80 and RDL40, independent of the anatomic level (p < 0.001). CNRD showed no significant differences at the abdominal and pelvic level between the investigated radiation dose levels. However, for thigh to foot level a significant increase of CNRD was noted between RDL120, RDL80 and RDL40. Significant differences of qualitative IQ were observed between RDL120 and RDL40 from the abdominal to the foot level, whereas no difference was seen for the other dose levels.

CONCLUSION: Radiation dose splitting with VSS-CT can be applied to run-off CTA facilitating intra-individual comparison of different acquisition protocols without additional radiation exposure. Furthermore, a radiation dose reduction potential for run-off CTA of approximately 1/3 as compared to the acquisition protocol recommended by the manufacturer could be identified in this study.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Werncke T, Hinrichs JB, Alikhani B, et al. Virtual single source CT using dual source acquisition: Clinical applicability in run-off CT-angiography for intra-individual comparison of different scan protocols. Eur J Radiol. 2018. PMID: 29571789. doi:10.1016/j.ejrad.2018.02.012

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。