先进模型迭代重建(ADMIRE)有助于腹部CT辐射剂量降低
本研究对11例连续腹部CT患者按本研究院标准方案(100 kV、289参考mAs)扫描,将图像重建为全剂量加权滤波反投影或模拟降至全剂量10%并联合ADMIRE强度3或5,由5名读者行二选一强制选择比较共3000次;ADMIRE 3与5均显著降低解剖结构噪声且两者间有显著差异,ADMIRE的辐射剂量降低潜力为29.0%至53.5%,ADMIRE 3与5在对比度亚组内无显著差异。
要点
- ADMIRE 3与5较滤波反投影均显著降低解剖结构噪声,且ADMIRE 5较3噪声降低更显著。
- ADMIRE的辐射剂量降低潜力为29.0%至53.5%,ADMIRE 3与5在对比度亚组内无显著差异。
- 腹部CT采用ADMIRE强度3可能允许约30%的辐射剂量降低。
核心结论
- ADMIRE 5较ADMIRE 3允许显著更高的噪声降低潜力,但并未实现更高水平的辐射剂量降低。
- ADMIRE具有显著的辐射剂量降低潜力,且该潜力在一定程度上依赖于感兴趣结构的对比度。
- 腹部CT采用ADMIRE强度3可能允许约30%的辐射剂量降低。
研究设计与样本
- 研究类型为对连续患者CT图像进行模拟剂量降低与不同重建算法的图像质量比较研究。
- 样本为11例连续患者,按本研究院标准方案行腹部CT扫描(100 kV、289参考mAs)。
- 图像重建为全剂量加权滤波反投影或模拟辐射剂量降低至全剂量水平的10%并联合ADMIRE强度3或5。
- 由5名读者采用二选一强制选择比较评估图像质量,共完成3000次比较。
关键影像表现/征象
- ADMIRE 3与5较滤波反投影均显著降低解剖结构噪声,且ADMIRE 3与5之间差异显著。
- ADMIRE的辐射剂量降低潜力为29.0%至53.5%,ADMIRE 3与5在对比度亚组内无显著差异。
- ADMIRE 3的辐射剂量降低潜力在高、中、低对比度结构间差异显著,ADMIRE 5仅在高对比度与中对比度亚组间差异显著。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: This study aimed to determine the potential degree of radiation dose reduction achievable using Advanced Modeled Iterative Reconstruction (ADMIRE) in abdominal computed tomography (CT) while maintaining image quality. Moreover, this study compared differences in image noise reduction of this iterative algorithm with radiation dose reduction.
METHODS: Eleven consecutive patients scheduled for abdominal CT were scanned according to our institute's standard protocol (100 kV, 289 reference mAs). Using a proprietary reconstruction software, CT images of these patients were reconstructed as either full-dose weighted filtered back projections or with simulated radiation dose reductions down to 10% of the full-dose level and ADMIRE at either strength 3 or strength 5. Images were marked with arrows pointing on anatomic structures of the abdomen, differing in their contrast to the surrounding tissue. Structures were grouped into high-, medium-, and low-contrast subgroups. In addition, the intrinsic noise of these structures was measured. That followed, image pairs were presented to observers, with five readers assessing image quality using two-alternative-forced-choice comparisons. In total, 3000 comparisons were performed that way.
RESULTS: Both ADMIRE 3 and 5 decreased noise of the anatomic structures significantly compared to the filtered back projection, with an additional significant difference between ADMIRE 3 and 5. Radiation dose reduction potential for ADMIRE ranged from 29.0% to 53.5%, with no significant differences between ADMIRE 3 and 5 within the contrast subgroups.The potential levels of radiation dose reduction for ADMIRE 3 differed significantly between high-, medium-, and low-contrast structures, whereas for ADMIRE 5, there was only a significant difference between the high- and the medium-contrast subgroups.
CONCLUSION: Although ADMIRE 5 permits significantly higher noise reduction potential than ADMIRE 3, it does not facilitate higher levels of radiation dose reduction. ADMIRE nonetheless holds remarkable potential for radiation dose reduction, which features a certain dependency on the contrast of the structure of interest. Applying ADMIRE with a strength of 3 in abdominal CT may permit radiation dose reduction of about 30%.
引用
- Ellmann S, Kammerer F, Allmendinger T, et al. Advanced Modeled Iterative Reconstruction (ADMIRE) Facilitates Radiation Dose Reduction in Abdominal CT. Acad Radiol. 2018. PMID: 29500115. doi:10.1016/j.acra.2018.01.014