从单张数字X线摄影图像确定儿科患者厚度:一项原理验证

Determining paediatric patient thickness from a single digital radiograph-a proof of principle.
PubMed 文献 · Br J Radiol · 2018-04-05

本研究提出并验证一种仅用单张X线摄影图像、曝光参数及X线设备与探测器先验知识估算衰减体厚度的方法。用11张已知厚度Solid Water体模图像在60、70和81 kVp下测试,估算厚度与已知厚度平均相差3.8 mm(3.2%),范围为0.5-10.8 mm(0.5-9%)。该方法有望成为估算儿科腹部患者厚度以辅助患者剂量审计的临床工具。

要点

  • 仅用单张X线摄影图像、曝光参数和X线设备与探测器先验知识即可估算均匀衰减体厚度。
  • 在60、70和81 kVp下对11张已知厚度Solid Water体模图像测试,估算厚度与已知厚度平均相差3.8 mm(3.2%)。
  • 该方法有望用于估算儿科腹部患者厚度以辅助患者剂量审计。

核心结论

  • 提出一种仅用单张X线摄影图像估算均匀衰减体厚度的原理验证方法,在Solid Water体模上显示准确,估算与已知厚度最大相差10.8 mm(9%),有望成为估算儿科腹部患者厚度以辅助患者剂量审计的临床工具。

研究设计与样本

  • 原理验证研究,使用Solid Water体模,在60、70和81 kVp下采集11张已知且不同厚度的测试图像,以已知厚度为参考进行比较。

诊断标准与关键数值

  • 估算厚度与已知厚度平均相差3.8 mm(3.2%),范围为0.5-10.8 mm(0.5-9%)。
  • 估算与已知厚度最大相差10.8 mm(9%)。

临床应用与阅片要点

  • 该方法仅依赖医学物理部门常规收集的数据类型,有望用于估算儿科腹部患者厚度以辅助患者剂量审计。

局限性

  • 本工作仅为原理验证,使用Solid Water体模而非患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This work presents a proof of principle for a method of estimating the thickness of an attenuator from a single radiograph using the image, the exposure factors with which it was acquired and a priori knowledge of the characteristics of the X-ray unit and detector used for the exposure. It is intended this could be developed into a clinical tool to assist with paediatric patient dose audit, for which a measurement of patient size is required.

METHODS: The proof of principle used measured pixel value and effective linear attenuation coefficient to estimate the thickness of a Solid Water attenuator. The kerma at the detector was estimated using a measurement of pixel value on the image and measured detector calibrations. The initial kerma was estimated using a lookup table of measured output values. The effective linear attenuation coefficient was measured for Solid Water at varying kVp. 11 test images of known and varying thicknesses of Solid Water were acquired at 60, 70 and 81 kVp. Estimates of attenuator thickness were made using the model and the results compared to the known thickness.

RESULTS: Estimates of attenuator thickness made using the model differed from the known thickness by 3.8 mm (3.2%) on average, with a range of 0.5-10.8 mm (0.5-9%).

CONCLUSION: A proof of principle is presented for a method of estimating the thickness of an attenuator using a single radiograph of the attenuator. The method has been shown to be accurate using a Solid Water attenuator, with a maximum difference between estimated and known attenuator thickness of 10.8 mm (9%). The method shows promise as a clinical tool for estimating abdominal paediatric patient thickness for paediatric patient dose audit, and is only contingent on the type of data routinely collected by Medical Physics departments. Advances in knowledge: A computational model has been created that is capable of accurately estimating the thickness of a uniform attenuator using only the radiographic image, the exposure factors with which it was acquired and a priori knowledge of the characteristics of the X-ray unit and detector used for the exposure.

来源:Br J Radiol,查看原文。版权归原作者所有。