减少肝脏点剪切波弹性成像测量次数:临床实践中影响肝纤维化评估测量次数及可靠性的因素
在232名参与者中获取10次点剪切波弹性成像肝脏剪切波速度测量,发现与10次测量相比最少需6次测量即可保持诊断性能;可靠性指标IQR/中位数比值≤30%时诊断不一致率低于>30%时,显著纤维化是低可靠性的独立预测因素(HR 2.22,P<.020)。
要点
- 点剪切波弹性成像评估肝纤维化时,最少6次剪切波速度测量即可替代10次测量。
- 6次与10次测量诊断显著纤维化和严重纤维化或肝硬化的AUC无显著差异。
- IQR/中位数比值≤30%的高可靠性测量诊断不一致率低于>30%的低可靠性测量。
核心结论
- 采用点剪切波弹性成像评估肝纤维化时,最少需6次剪切波速度测量(中位数6次 vs 中位数10次);测量次数对诊断准确性的影响小于低可靠性测量的影响。
研究设计与样本
- 研究纳入232名参与者,每人均获取10次点剪切波弹性成像肝脏剪切波速度测量,以Ishak分期≥3为显著纤维化、≥5为严重纤维化或肝硬化,比较前2至前9次测量中位数与全部10次测量中位数的组内相关系数,并比较6次与10次测量的诊断性能。
诊断标准与关键数值
- 与全部10次测量相比,组内相关系数大于0.95所需的最少测量次数为6次;6次与10次测量诊断显著纤维化的AUC分别为0.828与0.839(P=.487),诊断严重纤维化或肝硬化的AUC分别为0.953与0.969(P=.145)。
临床应用与阅片要点
- 采用10次测量时,高可靠性测量(IQR/中位数比值≤30%)与显著纤维化和严重纤维化或肝硬化之间的不一致率分别为150例中22例(14.7%)和150例中3例(2.0%),低可靠性测量(IQR/中位数比值>30%)分别为82例中26例(31.7%)和82例中8例(9.8%);显著纤维化是低可靠性的独立预测因素(HR 2.22;P<.020)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To identify the minimum number of measurements required for the noninvasive assessment of liver fibrosis by using point shear-wave elastography (pSWE) and determine whether the use of a reliability indicator such as interquartile range [IQR]-to-median ratio will affect diagnostic performance. Materials and Methods Ten liver shear-wave velocity (SWV) measurements by pSWE were obtained in 232 participants. Interclass correlation coefficients (ICC) between the median of the first two through the first nine measurements and all 10 measurements were calculated; the minimum number of measurements with ICC greater than 0.95 versus all 10 measurements was determined. The diagnostic performance of the minimum number of measurements and 10 measurements in identifying significant (Ishak stage, ≥3) and severe fibrosis or cirrhosis (Ishak stage, ≥5) was compared by using areas under the receiver operating characteristic curve. These were compared between measurements that demonstrated higher or lower reliability (IQR-to-median ratio of ≤ 30% and IQR-to-median ratio of > 30%, respectively). Results Compared with 10 measurements, a minimum of six SWV measurements was required. The overall area under the curve for diagnosing significant (areas under the receiver operating characteristic curve, 0.828 vs 0.839; P = .487) and severe fibrosis or cirrhosis (0.953 vs 0.969, respectively; P = .145) did not differ according to number of measurements (six vs 10); a median of six measurements resulted in only limited disagreement (nine of 232 [3.9%]) versus histologic evaluation. When using 10 measurements, higher reliability measurements showed a lower percentage of discordance between pSWE and significant fibrosis and severe fibrosis or cirrhosis (22 [14.7%] and three [2.0%] of 150 cases, respectively) compared with lower reliability measurements (26 [31.7%] and eight [9.8%] of 82 cases, respectively). Significant fibrosis was an independent predictor for lower reliability (hazard ratio, 2.22; P < .020). Conclusion A limited number of SWV measurements (median six vs median 10) were required for the assessment of liver fibrosis by using pSWE. The number of measurements had less influence on the diagnostic accuracy compared with lower reliability measurements. © RSNA, 2018 Online supplemental material is available for this article.
引用
- Fang C, Jaffer OS, Yusuf GT, et al. Reducing the Number of Measurements in Liver Point Shear-Wave Elastography: Factors that Influence the Number and Reliability of Measurements in Assessment of Liver Fibrosis in Clinical Practice. Radiology. 2018. PMID: 29514018. doi:10.1148/radiol.2018172104
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。