经腔内途径联合气泡种植组织碎化术利用超声机械效应治疗癌症

Transluminal Approach with Bubble-Seeded Histotripsy for Cancer Treatment with Ultrasonic Mechanical Effects.
PubMed 文献 · Ultrasound Med Biol · 2018-03-07

本研究开发了一种可附着于模拟超声内镜探头的20×20 mm方形小孔径、35 mm长焦距、2.1 MHz换能器,结合局部注射相变纳米液滴与脉冲高强度聚焦超声,在离体和在体研究中以约4 kW/cm²的声强实现了可预测的组织机械碎化,该声强在不使用相变纳米液滴时不足以诱导显著生物效应。利用超快超声成像监测微泡运动,发现脉冲高强度聚焦超声暴露后25-30秒微泡运动最大,40秒后降至初始水平,提示过度暴露因热效应导致凝固。

要点

  • 经腔内途径联合气泡种植组织碎化术使用小孔径长焦距换能器,在约4 kW/cm²声强下实现组织机械碎化。
  • 超快超声成像可实时监测微泡运动,25-30秒运动最大,40秒后降至初始水平提示过度暴露致热凝固。
  • 不使用相变纳米液滴时,相同声强不足以诱导显著生物效应。

核心结论

  • 经腔内途径联合气泡种植组织碎化术使用小孔径换能器在实时监测下可实现可预测的组织机械碎化,有望扩展该技术对深部组织的应用。

研究设计与样本

  • 初步研究,开发小孔径(20×20 mm方形)、长焦距(35 mm)、2.1 MHz换能器,可附着于模拟超声内镜探头,结合局部注射相变纳米液滴与脉冲高强度聚焦超声,在离体和在体研究中评估组织机械碎化,并使用超快超声成像监测微泡运动。

诊断标准与关键数值

  • 使用相变纳米液滴时,在约4 kW/cm²声强下观察到可预测的组织机械碎化;不使用相变纳米液滴时,相同声强不足以诱导显著生物效应。
  • 脉冲高强度聚焦超声暴露后25-30秒微泡运动最大,40秒后微泡运动降至与操作开始时相同水平。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Bubble-seeded histotripsy (BSH) is a newly developed ultrasound-based mechanical fractionation technique using locally injected phase change nanodroplets (PCNDs) as sensitizers. The PCNDs are a kind of microbubble precursor compressed into submicron-size in droplets form, which were designed for local administration and will expand into microbubbles under ultrasound exposure. Previously, we reported that a combination of PCNDs injection and pulsed high-intensity focused ultrasound (pHIFU) with an acoustic intensity as low as about 3 kW/cm2 at 1.1 MHz, which is similar to the acoustic intensity of currently available HIFU coagulation therapy, was enough to induce tissue fractionation after significant antitumor effects in an in vivo study. Toward therapeutic application of BSH to deep-seated tissues such as the pancreas, the transluminal approach, using endoscopic ultrasound was thought to be ideal. Therefore, for a preliminary examination, we developed a new transducer with a small aperture (20- × 20-mm square) and long focal length (35 mm), operating at 2.1 MHz that could be attached to an EUS-mimicking probe. With the newly developed transducer and locally injected PCNDs, predictable tissue mechanical fractionation was observed in both ex vivo and in vivo studies at acoustic intensities that were too low to induce any significant bioeffects (around 4 kW/cm2) without using PCNDs. For in situ monitoring of the treatment site during a procedure, the degree of attenuation of microbubble motions after exposing the microbubbles to pHIFU was monitored, using ultrafast echographic imaging. Microbubble movements were observed to be largest at 25-30 s after pHIFU exposure. On the contrary, after 40 s, the movement of microbubbles decreased to the same level as at the start of the procedure, suggesting that an overdose of pHIFU exposure causes coagulation attributable to the thermal effect caused by absorption of the energy. Those results were promising for expanding the application of BSH for a transluminal approach, using a small transducer under real-time monitoring.