文献速递腹部泌尿生殖技术与安全CT

第三代双源双能CT虚拟平扫与真实平扫腹部衰减值的比较

Abdominal Attenuation Values on Virtual and True Unenhanced Images Obtained With Third-Generation Dual-Source Dual-Energy CT.
PubMed 文献 · AJR Am J Roentgenol · 2018-03-12

这项前瞻性研究纳入83例需胸腹部CT肿瘤检查的患者,比较第三代双源双能CT虚拟平扫与真实平扫的腹部组织衰减值差异。除门静脉期肝脾及两期肌肉外,各组织衰减值均有显著差异;VNC1图像上肾脏约30%、胆囊和膀胱约70%、脂肪40-70%的病例差异≥10 HU,VNC2图像上脂肪约90%的病例差异≥20 HU。结论认为腹部虚拟平扫仍不应替代真实平扫。

要点

  • 除门静脉期肝脾及两期肌肉外,虚拟平扫与真实平扫各组织衰减值均有显著差异(p=<0.001-0.042)。
  • VNC1图像上肾脏约30%、胆囊和膀胱约70%、脂肪40-70%的病例差异≥10 HU。
  • VNC2图像上脂肪约90%的病例差异≥20 HU,腹部虚拟平扫仍不应替代真实平扫。

核心结论

  • 第三代双源双能CT腹部虚拟平扫在液体、脂肪和肾组织衰减值上与真实平扫仍有实质差异,仍不应替代真实平扫。

研究设计与样本

  • 前瞻性研究,纳入83例需胸腹部CT行肿瘤检查的患者,行真实平扫(管电压120 kVp)及动脉期和门静脉期双能CT(管电压100 kVp与Sn 150 kVp,管电流-曝光时间乘积190与95 mAs),用两种商用双能CT后处理算法生成虚拟平扫图像(VNC1为肝脏设计,VNC2为含少量脂肪器官设计)。

诊断标准与关键数值

  • 测量肝脏、脾脏、椎旁肌肉、腹膜后脂肪、肾皮质和髓质、胆囊和膀胱腔的衰减值;差异≥10 HU的病例比例在VNC1图像上肾脏约30%、胆囊和膀胱约70%、脂肪依期相为40-70%,VNC2图像上脂肪差异≥20 HU的病例约90%。

临床应用与阅片要点

  • 腹部虚拟平扫在液体、脂肪和肾组织衰减值上与真实平扫差异较大,阅片和报告时需注意这些组织衰减值可能不可靠。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study is to investigate the magnitude of differences between attenuation values measured on virtual unenhanced images and true unenhanced images obtained using third-generation dual-source dual-energy CT (DECT).

SUBJECTS AND METHODS: A total of 83 patients requiring thoracoabdominal CT for cancer workup were included in this prospective study. CT examinations included true unenhanced acquisitions (tube potential, 120 kVp) and arterial and portal phase dual-energy CT (DECT) acquisitions (tube potential, 100 kVp and Sn 150 kVp [where Sn denotes the interposition of a tin filter in the high-energy beam]; tube current-exposure time product, 190 and 95 mAs). Virtual unenhanced images were created using two commercially available DECT postprocessing algorithms, one of which was designed to create liver images (hereafter referred to as VNC1 images) and the other of which was designed to create images of organs containing minor amounts of fat (hereafter referred to as VNC2 images). Attenuation values on the liver, spleen, paraspinal muscles, retroperitoneal fat, renal cortex and medulla, and gallbladder and bladder lumens were measured.

RESULTS: The attenuation values of all tissues were significantly different between virtual unenhanced and true unenhanced images (p = < 0.001-0.042), except for the liver and spleen in the portal phase and muscles in both phases. When statistically significant, correlations between these differences and body mass index (weight in kilograms divided by the square of height in meters) depended on the tissue imaged and algorithm used. The percentage of cases in which these differences were 10 HU or greater was 1% for the liver and approximately 5% for the spleen and muscles, regardless of the algorithm and phase, but on VNC1 images it reached approximately 30% for the kidney, 70% for the gallbladder and bladder, and depending on the phase, 40-70% for fat. On VNC2 images, the percentage of cases in which these differences were 20 HU or greater was approximately 90% for fat.

CONCLUSION: Abdominal virtual unenhanced images obtained with third-generation dual-source DECT still should not replace true unenhanced images because of substantial differences in attenuation measurements for fluid, fat, and renal tissues.