超高清CT冠状动脉CT血管造影的诊断效能:与有创冠状动脉造影的比较
回顾性纳入38例疑似冠心病患者,行超高清CT冠状动脉CT血管造影(CCTA)后以有创冠状动脉造影(ICA)为参考标准。患者、血管、节段水平敏感度分别为100%、96%、95%,特异度分别为67%、81%、96%;CCTA直径狭窄百分比与ICA相关性R=0.90,平均高估4%±7%(p<.01)。结论:超高清CT CCTA在冠状动脉狭窄定量方面与ICA相关性和一致性极好。
要点
- 超高清CT CCTA对阻塞性冠心病患者水平敏感度100%(95%CI 95%-100%),特异度67%(95%CI 38%-67%)。
- CCTA直径狭窄百分比与ICA相关性R=0.90(95%CI 0.83-0.95),平均高估4%±7%(p<.01),一致性界限±16%。
- CCTA中位有效辐射剂量5.4 mSv(范围2.9-18.0 mSv)。
核心结论
- 超高清CT冠状动脉CT血管造影在冠状动脉狭窄定量方面与有创冠状动脉造影显示极好的相关性和一致性。
研究设计与样本
- 回顾性研究,纳入38例连续疑似冠心病患者,均行超高清CT(0.25 mm×128排探测器)冠状动脉CT血管造影后行有创冠状动脉造影,以有创冠状动脉造影为参考标准。
诊断标准与关键数值
- 患者水平:敏感度100%(95%CI 95%-100%),特异度67%(95%CI 38%-67%)。
- 血管水平:敏感度96%(95%CI 89%-99%),特异度81%(95%CI 75%-83%)。
- 节段水平:敏感度95%(95%CI 89%-98%),特异度96%(95%CI 96%-97%)。
- 直径狭窄>30%的节段:CCTA与ICA直径狭窄百分比相关性R=0.90(95%CI 0.83-0.95),平均高估4%±7%(p<.01),一致性界限±16%。
- ICA显示阻塞性冠心病见于32例患者(84%)的51支血管(45%)的65个节段(12%)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Recently, ultra-high-resolution computed tomography (U-HRCT) with a 0.25 mm × 128-row detector was introduced. The purpose of this study was to evaluate the diagnostic performance of coronary CT angiography (CCTA) using U-HRCT.
METHODS: This retrospective study included 38 consecutive patients with suspected coronary artery disease (CAD) who underwent CCTA with U-HRCT followed by invasive coronary angiography (ICA). Per-segment diameter stenosis was calculated. Diagnostic performance of CCTA relative to ICA as the reference standard was determined. For segments with >30% diameter stenosis, the correlation and agreement of percent diameter stenosis between CCTA and ICA were calculated.
RESULTS: Obstructive CAD was observed in 65 segments (12%) of 51 vessels (45%) in 32 patients (84%) during ICA. The per-patient, vessel, and segment analyses showed a sensitivity of 100% (95% confidence interval [CI], 95%-100%), 96% (95% CI: 89%-99%) and 95% (95% CI: 89%-98%), respectively, and a specificity of 67% (95% CI: 38%-67%), 81% (95% CI: 75%-83%) and 96% (95% CI: 96%-97%), respectively. The percentage of diameter stenosis, as determined by CCTA, demonstrated an excellent correlation with ICA (R = 0.90; 95% CI: 0.83-0.95) and a slight significant overestimation (mean: 4% ± 7%, p < .01), with the agreed range of limits being ± 16%. The median effective radiation dose for CCTA was 5.4 mSv (range: 2.9-18.0 mSv).
CONCLUSIONS: CCTA with U-HRCT demonstrated an excellent correlation and agreement with ICA in the quantification of coronary artery stenosis.
引用
- Takagi H, Tanaka R, Nagata K, et al. Diagnostic performance of coronary CT angiography with ultra-high-resolution CT: Comparison with invasive coronary angiography. Eur J Radiol. 2018. PMID: 29571798. doi:10.1016/j.ejrad.2018.01.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。