文献速递外周血管技术与安全CT

实施基于体型的CT血管造影本地诊断参考水平

Implementation of Size-Dependent Local Diagnostic Reference Levels for CT Angiography.
PubMed 文献 · AJR Am J Roentgenol · 2018-03-23

本研究纳入184例主动脉CTA检查(第二代或第三代双源CT),提出按体型设定本地诊断参考水平(LDRL)的方法,并与体型无关LDRL比较及评估图像质量。结果显示CTDIvol取决于患者体型和扫描仪类型,体型相关LDRL对评估个体患者辐射剂量适宜性有相关性,所有离群值主观图像质量评分均为足够或更好。

要点

  • CTDIvol与患者体型和扫描仪类型相关(R²=0.72和0.78,斜率=0.05和0.02 mGy/mm,p<0.001)。
  • 体型无关LDRL识别的离群值中,30%(CT1)和67%(CT2)在考虑患者体型时剂量适当。
  • 体型相关LDRL识别的离群值中,30%(CT1)和70%(CT2)未被体型无关LDRL识别。

核心结论

  • 研究提出了一种在CTA中设定体型相关LDRL的方法。
  • 体型相关LDRL对于评估个体患者在特定CT扫描仪上的辐射剂量适宜性有相关性。

研究设计与样本

  • 研究纳入184例主动脉CTA检查,使用第二代双源CT(CT1)或第三代双源CT(CT2)扫描仪。
  • 通过剂量监测软件获取容积CT剂量指数(CTDIvol)和患者直径(水当量直径)。
  • 按扫描仪类型基于CTDIvol与患者体型的线性回归设定体型相关LDRL。
  • 体型无关LDRL设定为CTDIvol值的第5和第95百分位数。
  • 客观图像质量采用信噪比(SNR)评估,主观图像质量采用4点Likert量表评估。

诊断标准与关键数值

  • CTDIvol与患者体型和扫描仪类型相关(R²=0.72和0.78,斜率=0.05和0.02 mGy/mm,p<0.001)。
  • 体型无关LDRL识别的离群值中,30%(CT1)和67%(CT2)在考虑患者体型时剂量适当。
  • 体型相关LDRL识别的离群值中,30%(CT1)和70%(CT2)未被体型无关LDRL识别。
  • SNR与CTDIvol呈负相关(R²=0.36 for CT1和0.45 for CT2)。
  • 所有离群值的主观图像质量评分均为足够或更好。

临床应用与阅片要点

  • 体型相关LDRL可用于评估个体患者在特定CT扫描仪上的辐射剂量适宜性。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Diagnostic reference levels (DRLs) are established for standard-sized patients; however, patient dose in CT depends on patient size. The purpose of this study was to introduce a method for setting size-dependent local diagnostic reference levels (LDRLs) and to evaluate these LDRLs in comparison with size-independent LDRLs and with respect to image quality.

MATERIALS AND METHODS: One hundred eighty-four aortic CT angiography (CTA) examinations performed on either a second-generation or third-generation dual-source CT scanner were included; we refer to the second-generation dual-source CT scanner as "CT1" and the third-generation dual-source CT scanner as "CT2." The volume CT dose index (CTDIvol) and patient diameter (i.e., the water-equivalent diameter) were retrieved by dose-monitoring software. Size-dependent DRLs based on a linear regression of the CTDIvol versus patient size were set by scanner type. Size-independent DRLs were set by the 5th and 95th percentiles of the CTDIvol values. Objective image quality was assessed using the signal-to-noise ratio (SNR), and subjective image quality was assessed using a 4-point Likert scale.

RESULTS: The CTDIvol depended on patient size and scanner type (R2 = 0.72 and 0.78, respectively; slope = 0.05 and 0.02 mGy/mm; p < 0.001). Of the outliers identified by size-independent DRLs, 30% (CT1) and 67% (CT2) were adequately dosed when considering patient size. Alternatively, 30% (CT1) and 70% (CT2) of the outliers found with size-dependent DRLs were not identified using size-independent DRLs. A negative correlation was found between SNR and CTDIvol (R2 = 0.36 for CT1 and 0.45 for CT2). However, all outliers had a subjective image quality score of sufficient or better.

CONCLUSION: We introduce a method for setting size-dependent LDRLs in CTA. Size-dependent LDRLs are relevant for assessing the appropriateness of the radiation dose for an individual patient on a specific CT scanner.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Boere H, Eijsvoogel NG, Sailer AM, et al. Implementation of Size-Dependent Local Diagnostic Reference Levels for CT Angiography. AJR Am J Roentgenol. 2018. PMID: 29570369. doi:10.2214/AJR.17.18566

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。