双能冠状动脉CT血管成像对比剂注射方案优化:循环体模研究结果
在循环体模上使用第三代双源CT研究双能冠状动脉CTA达到诊断性冠状动脉衰减(300 HU)所需的最低碘流率。碘流率从1.0至0.3 gI/s以0.1-gI/s间隔变化,40-70 keV的VMI+重建在至少一个碘流率下达到诊断性衰减,最低碘流率为40 keV下的0.4 gI/s(312.8 HU)。
要点
- 40-70 keV的VMI+重建可在至少一个碘流率下达到诊断性冠状动脉衰减。
- 达到诊断性衰减的最低碘流率从40 keV的0.4 gI/s(312.8 HU)至70 keV的1.0 gI/s(334.1 HU)。
- CNR与碘流率呈近完美相关(ρ≥0.962;p<0.001),1.0 gI/s在各keV水平提供最高CNR,40 keV时最高(54.0±3.1)。
核心结论
- 与40-70 keV的VMI+重建联合时,0.4-1.0 gI/s的碘流率可在双能冠状动脉CTA中提供诊断性冠状动脉衰减。
研究设计与样本
- 在循环体模上使用第三代双源CT扫描仪进行采集;对比剂固定时间注射,碘流率从1.0至0.3 gI/s以0.1-gI/s间隔变化;生成40至90 keV、以5 keV递增的噪声优化虚拟单能成像(VMI+)重建。
关键影像表现/征象
- 40-70 keV的VMI+重建在至少一个碘流率下达到诊断性衰减。
- 达到诊断性衰减的最低碘流率从40 keV的0.4 gI/s(312.8 HU)至70 keV的1.0 gI/s(334.1 HU)。
- 碘流率1.0 gI/s在各keV水平达到的衰减值显著更高(p<0.001)。
- CNR与碘流率呈近完美相关(ρ≥0.962;p<0.001)。
- 碘流率1.0 gI/s在各keV水平提供最高CNR,40 keV时达最高总体值(54.0±3.1)。
诊断标准与关键数值
- 诊断性冠状动脉衰减定义为300 HU。
- 最低碘流率达到诊断性衰减:40 keV下0.4 gI/s(312.8 HU)至70 keV下1.0 gI/s(334.1 HU)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the minimum iodine delivery rate (IDR) required to achieve diagnostic coronary attenuation (300 HU) with dual-energy coronary CTA.
METHODS: Acquisitions were performed on a circulation phantom with a third- generation dual-source CT scanner. Contrast media was injected for a fixed time whilst IDRs varied from 1.0 to 0.3 gI/s in 0.1-gI/s intervals. Noise-optimized virtual monoenergetic imaging (VMI+) reconstructions from 40 to 90 keV in 5 keV increments were generated. Contrast-to-noise ratio (CNR) and coronary HU were measured for each injection.
RESULTS: VMI+ from 40-70 keV reached diagnostic attenuation with at least one IDR. The minimum IDR achieving a diagnostic attenuation ranged from 0.4 gI/s at 40 keV (312.8 HU) to 1.0 gI/s at 70 keV (334.1 HU). Attenuation values reached with IDR of 1.0 gI/s were significantly higher at each keV level (p<0.001). CNR showed a near perfect correlation with the IDR (ρ≥0.962; p<0.001), the IDR of 1.0 gI/s provided the highest CNR at each keV level, achieving the highest overall value at 40 keV (54.0±3.1).
CONCLUSIONS: IDRs from 0.4-1.0 gI/s associated with VMI+ from 40-70 keV provide diagnostic coronary attenuation with dual-energy coronary CTA.
KEY POINTS: • Iodine delivery rate (IDR) is a major determinant of contrast enhancement. • Low-keV noise-optimized monoenergetic images (VMI+) maximize iodine attenuation. • Low-keV VMI+ allows for lower IDRs while maintaining adequate coronary attenuation. • Lowest IDR to reach 300 HU was 0.4 gI/s, 40 keV VMI+.
引用
- De Santis D, Caruso D, Schoepf UJ, et al. Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom. Eur Radiol. 2018. PMID: 29488083. doi:10.1007/s00330-018-5308-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。