文献速递胸部CT

标准剂量、缩减剂量与超低剂量CT联合或不联合混合型迭代重建技术对肺结节计算机辅助检测(CADe)能力的比较

Comparison of computer-aided detection (CADe) capability for pulmonary nodules among standard-, reduced- and ultra-low-dose CTs with and without hybrid type iterative reconstruction technique.
PubMed 文献 · Eur J Radiol · 2018-01-11

本研究对40例患者行标准剂量(250 mA)、缩减剂量(50 mA)和超低剂量(10 mA)胸部CT,重建为1 mm层厚,比较联合或不联合自适应迭代剂量减低3D(AIDR 3D)时CADe软件检测≥4 mm结节的敏感度和假阳性率。超低剂量CT不加AIDR 3D的敏感度0.43、假阳性率7.88,显著劣于其他方案;AIDR 3D对CADe结节检测能力有显著正向影响,即使辐射剂量降低。

要点

  • 超低剂量CT不加AIDR 3D时CADe敏感度仅0.43,假阳性率7.88,显著劣于其他方案。
  • 标准剂量CT联合AIDR 3D时CADe敏感度0.78、假阳性率3.05。
  • AIDR 3D对CADe结节检测能力有显著正向影响,即使辐射剂量降低。

核心结论

  • AIDR 3D对CADe系统的结节检测能力有显著正向影响,即使辐射剂量降低。

研究设计与样本

  • 经机构审查委员会批准并获书面知情同意的前瞻性研究,40例患者(男21例,平均年龄63.1±11.0岁;女19例,平均年龄65.1±12.7岁)行标准剂量(250 mA)、缩减剂量(50 mA)和超低剂量(10 mA)胸部CT,重建为1 mm层厚,用专有CADe软件自动检测直径≥4 mm的结节,采用McNemar检验或符号秩检验比较各方案的敏感度和假阳性率(/例)。

诊断标准与关键数值

  • CADe检测结节直径阈值:≥4 mm。
  • 超低剂量CT不加AIDR 3D:敏感度0.43,假阳性率7.88。
  • 标准剂量CT加AIDR 3D:敏感度0.78,假阳性率3.05;不加AIDR 3D:敏感度0.80,假阳性率2.63。
  • 缩减剂量CT加AIDR 3D:敏感度0.81,假阳性率3.05;不加AIDR 3D:敏感度0.62,假阳性率2.95。
  • 超低剂量CT加AIDR 3D:敏感度0.79,假阳性率4.88。

临床应用与阅片要点

  • 原文结论为AIDR 3D对CADe系统的结节检测能力有显著正向影响,即使辐射剂量降低。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To directly compare the effect of a reconstruction algorithm on nodule detection capability of the computer-aided detection (CADe) system using standard-dose, reduced-dose and ultra-low dose chest CTs with and without adaptive iterative dose reduction 3D (AIDR 3D).

MATERIALS AND METHODS: Our institutional review board approved this study, and written informed consent was obtained from each patient. Standard-, reduced- and ultra-low-dose chest CTs (250 mA, 50 mA and 10 mA) were used to examine 40 patients, 21 males (mean age ± standard deviation: 63.1 ± 11.0 years) and 19 females (mean age, 65.1 ± 12.7 years), and reconstructed as 1 mm-thick sections. Detection of nodule equal to more than 4 mm in dimeter was automatically performed by our proprietary CADe software. The utility of iterative reconstruction method for improving nodule detection capability, sensitivity and false positive rate (/case) of the CADe system using all protocols were compared by means of McNemar's test or signed rank test.

RESULTS: Sensitivity (SE: 0.43) and false-positive rate (FPR: 7.88) of ultra-low-dose CT without AIDR 3D was significantly inferior to those of standard-dose CTs (with AIDR 3D: SE, 0.78, p < .0001, FPR, 3.05, p < .0001; and without AIDR 3D: SE, 0.80, p < .0001, FPR: 2.63, p < .0001), reduced-dose CTs (with AIDR 3D: SE, 0.81, p < .0001, FPR, 3.05, p < .0001; and without AIDR 3D: SE, 0.62, p < .0001, FPR: 2.95, p < .0001) and ultra-low-dose CT with AIDR 3D (SE, 0.79, p < .0001, FPR, 4.88, p = .0001).

CONCLUSION: The AIDR 3D has a significant positive effect on nodule detection capability of the CADe system even when radiation dose is reduced.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Ohno Y, Aoyagi K, Chen Q, et al. Comparison of computer-aided detection (CADe) capability for pulmonary nodules among standard-, reduced- and ultra-low-dose CTs with and without hybrid type iterative reconstruction technique. Eur J Radiol. 2018. PMID: 29496079. doi:10.1016/j.ejrad.2018.01.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。