文献速递技术与安全神经胸部CT

四分之一毫米光子计数计算机断层扫描的剂量效率:首次人体研究结果

Dose Efficiency of Quarter-Millimeter Photon-Counting Computed Tomography: First-in-Human Results.
PubMed 文献 · Invest Radiol · 2018-01-01

本研究在8名受试者中评估0.25毫米光子计数探测器CT的临床可行性、图像质量与辐射剂量影响。体模和犬类实验显示分辨率从9线对/厘米提升至18线对/厘米,但图像噪声增加75%;在相同平面分辨率和层厚下,0.25毫米模式噪声比标准分辨率低19%。

要点

  • 0.25毫米光子计数CT将空间分辨率从9线对/厘米提升至18线对/厘米。
  • 充分利用高分辨率时图像噪声较剂量匹配的0.5毫米标准分辨率采集增加75%。
  • 相同平面分辨率和层厚下,0.25毫米探测器模式噪声比标准分辨率低19%。

核心结论

  • 人体高分辨率光子计数CT在空间分辨率和图像噪声方面均优于标准分辨率光子计数CT。

研究设计与样本

  • 技术可行性研究,采用全身原型光子计数探测器CT扫描仪,先进行体模和犬类研究,随后8名人类受试者(平均年龄58±8岁,2名男性)接受脑、胸部及左肾上极轴位0.25毫米扫描。

诊断标准与关键数值

  • 线对体模中分辨率从9线对/厘米提升至18线对/厘米。
  • 调制传递函数在10%调制传递函数处从9.5线对/厘米提升至15.8线对/厘米。
  • 与剂量匹配的0.5毫米标准分辨率采集相比,充分利用高分辨率时图像噪声增加75%。
  • 与相同平面分辨率和层厚的标准分辨率数据相比,0.25毫米探测器模式在体模、动物和人体扫描中图像噪声低19%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of this study was to assess the clinical feasibility, image quality, and radiation dose implications of 0.25-mm imaging mode in a cohort of humans, achieved by dividing the photon-counting detector (PCD) size in half compared with standard-resolution photon-counting computed tomography (CT) (0.5 mm).

METHODS: In this technical feasibility study, a whole-body prototype PCD-CT scanner was studied in the 0.25 mm detector mode (measured at isocenter). A high-resolution PCD-CT protocol was first tested in phantom and canine studies in terms of image noise and spatial resolution. Then, 8 human subjects (mean age, 58 ± 8 years; 2 men) underwent axial PCD 0.25-mm scans of the brain, the thorax, and at the level of the upper left kidney. Filtered backprojection reconstruction was performed with a sharp kernel (B70) for standard-resolution and high-resolution data at 0.5-mm isotropic image voxel. High-resolution data, in addition, were reconstructed with an ultrasharp kernel (U70) at 0.25-mm isotropic voxels.

RESULTS: Image reconstructions from the PCD 0.25-mm detector system led to an improvement in resolution from 9 to 18 line pairs/cm in a line pair phantom. Modulation transfer function improved from 9.5 to 15.8 line pairs/cm at 10% modulation transfer function. When fully exploiting this improvement, image noise increased by 75% compared with dose-matched 0.5-mm slice PCD standard-resolution acquisition. However, when comparing with standard-resolution data at same in-plane resolution and slice thickness, the PCD 0.25-mm detector mode showed 19% less image noise in phantom, animal, and human scans.

CONCLUSION: High-resolution photon-counting CT in humans showed improved image quality in terms of spatial resolution and image noise compared with standard-resolution photon-counting.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Pourmorteza A, Symons R, Henning A, et al. Dose Efficiency of Quarter-Millimeter Photon-Counting Computed Tomography: First-in-Human Results. Invest Radiol. 2018. PMID: 29595753. doi:10.1097/RLI.0000000000000463

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。