文献速递神经外周血管CTDSA

模拟低剂量灌注CT检测动脉瘤性蛛网膜下腔出血后脑灌注损害的诊断准确性:一项回顾性分析

Diagnostic Accuracy of Simulated Low-Dose Perfusion CT to Detect Cerebral Perfusion Impairment after Aneurysmal Subarachnoid Hemorrhage: A Retrospective Analysis.
PubMed 文献 · Radiology · 2018-01-08

本回顾性研究纳入85例动脉瘤性蛛网膜下腔出血后怀疑脑血管痉挛的患者,比较模拟低剂量(72 mAs)与原始剂量(180 mAs)容积灌注CT检测脑灌注损害的诊断准确性。低剂量灌注CT图像质量与诊断信心良好,以原始灌注CT为标准,敏感度99.0%、特异度99.5%,两者与DSA的相关系数分别为r=0.667和r=0.671,病灶体积相对差异5.9%±5.1(r=0.955)。

要点

  • 模拟低剂量(72 mAs)灌注CT检测灌注损害的敏感度99.0%、特异度99.5%(以原始灌注CT为标准)。
  • 低剂量与原始剂量灌注CT对灌注损害有无及严重程度的评估无显著差异(Z=-0.447;P=.655)。
  • 低剂量灌注CT病灶体积与原始剂量相对差异5.9%±5.1,相关性r=0.955。

核心结论

  • 结果表明,在动脉瘤性蛛网膜下腔出血患者的容积灌注CT成像中,将辐射剂量降至原始水平的40%(管电流-时间乘积72 mAs)可能不会影响检测提示血管痉挛的脑灌注损害的诊断准确性。

研究设计与样本

  • 回顾性研究,纳入85例(平均年龄59.6岁;62例女性)动脉瘤性蛛网膜下腔出血且平扫CT和容积灌注CT怀疑脑血管痉挛的患者,其中37例在6小时内接受了数字减影血管造影(DSA);通过经验证的真实模拟回顾性生成管电流-时间乘积72 mAs的低剂量容积灌注CT数据集;灌注图由两名神经放射科医生评估,一名介入神经放射科医生评估DSA的16个血管节段。

诊断标准与关键数值

  • 以原始容积灌注CT为标准,低剂量容积灌注CT分类所有患者均正确,敏感度99.0%、特异度99.5%;两种数据集与DSA的相关系数分别为原始r=0.671、低剂量r=0.667。

临床应用与阅片要点

  • 作者认为,在动脉瘤性蛛网膜下腔出血患者的容积灌注CT成像中,将辐射剂量降至原始水平的40%(管电流-时间乘积72 mAs)可能不会影响检测提示血管痉挛的脑灌注损害的诊断准确性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To evaluate diagnostic accuracy of low-dose volume perfusion (VP) computed tomography (CT) compared with original VP CT regarding the detection of cerebral perfusion impairment after aneurysmal subarachnoid hemorrhage. Materials and Methods In this retrospective study, 85 patients (mean age, 59.6 years; 62 women) with aneurysmal subarachnoid hemorrhage and who were suspected of having cerebral vasospasm at unenhanced CT and VP CT (tube voltage, 80 kVp; tube current-time product, 180 mAs) were included, 37 of whom underwent digital subtraction angiography (DSA) within 6 hours. Low-dose VP CT data sets at tube current-time product of 72 mAs were retrospectively generated by validated realistic simulation. Perfusion maps were generated from both data sets and reviewed by two neuroradiologists for overall image quality, diagnostic confidence and presence and/or severity of perfusion impairment indicating vasospasm. An interventional neuroradiologist evaluated 16 vascular segments at DSA. Diagnostic accuracy of low-dose VP CT was calculated with original VP CT as reference standard. Agreement between findings of both data sets was assessed by using weighted Cohen κ and findings were correlated with DSA by using Spearman correlation. After quantitative volumetric analysis, lesion volumes were compared on both VP CT data sets. Results Low-dose VP CT yielded good ratings of image quality and diagnostic confidence and classified all patients correctly with high diagnostic accuracy (sensitivity, 99.0%; specificity, 99.5%) without significant differences regarding presence and/or severity of perfusion impairment between original and low-dose data sets (Z = -0.447; P = .655). Findings of both data sets correlated significantly with DSA (original, r = 0.671; low dose, r = 0.667). Lesion volume was comparable for both data sets (relative difference, 5.9% ± 5.1 [range, 0.2%-25.0%; median, 4.0%]) with strong correlation (r = 0.955). Conclusion The results suggest that radiation dose reduction to 40% of original dose levels (tube current-time product, 72 mAs) may be performed in VP CT imaging of patients with aneurysmal subarachnoid hemorrhage without compromising the diagnostic accuracy regarding detection of cerebral perfusion impairment indicating vasospasm. © RSNA, 2018 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Afat S, Brockmann C, Nikoubashman O, et al. Diagnostic Accuracy of Simulated Low-Dose Perfusion CT to Detect Cerebral Perfusion Impairment after Aneurysmal Subarachnoid Hemorrhage: A Retrospective Analysis. Radiology. 2018. PMID: 29309735. doi:10.1148/radiol.2017162707

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。