文献速递胸部心血管儿科影像CT

儿童与成人电门控双源心胸CT胸痛方案比较:管电流饱和对辐射剂量降低的影响

Comparison of Chest Pain Protocols for Electrocardiography-Gated Dual-Source Cardiothoracic CT in Children and Adults: The Effect of Tube Current Saturation on Radiation Dose Reduction.
PubMed 文献 · Korean J Radiol · 2018-01-02

本研究对比104例儿童与成人(16.6±7.7岁,范围5-48岁)在双源回顾性心电门控心胸CT中常规方案与胸痛方案的辐射剂量,并评估管电流饱和的影响;双相胸痛方案的容积CT剂量指数显著低于常规和单相方案,但管电流饱和时剂量节省效果显著低于未饱和时(4.5±10.2% vs 14.8±11.5%,p<0.001)。

要点

  • 双相胸痛方案的容积CT剂量指数显著低于常规和单相胸痛方案。
  • 管电流饱和时双相方案剂量节省效果显著低于未饱和时。
  • 在100 kVp的76例患者中,管电流饱和组容积CT剂量指数和每旋转管电流时间乘积显著更高。

核心结论

  • 在儿童和成人中,若进一步优化辐射剂量参数以避免管电流饱和,使用双相胸痛方案替代常规方案可使双源回顾性心电门控螺旋心胸CT的辐射剂量降低约15%。

研究设计与样本

  • 研究纳入104例患者(16.6±7.7岁,范围5-48岁),分为管电流饱和组与未饱和组;采用配对t检验比较同一患者相同成像参数下常规、单相和双相胸痛方案的估计辐射剂量,采用非配对t检验比较两组间的剂量降低百分比、患者年龄、容积CT剂量指数值和每旋转管电流时间乘积。

诊断标准与关键数值

  • 双相胸痛方案的容积CT剂量指数为10.8±3.9 mGy,显著低于常规方案的12.2±4.7 mGy(p<0.001)。
  • 双相胸痛方案的容积CT剂量指数为10.8±3.9 mGy,显著低于单相胸痛方案的12.9±4.9 mGy(p<0.001)。
  • 管电流饱和时双相胸痛方案的剂量节省效果为4.5±10.2%,显著低于未饱和时的14.8±11.5%(p<0.001)。
  • 在100 kVp的76例患者中,各方案管电流饱和组的容积CT剂量指数显著高于未饱和组(p<0.01),每旋转管电流时间乘积也显著更高(p<0.001)。

临床应用与阅片要点

  • 使用双相胸痛方案替代常规方案可降低辐射剂量,但需优化辐射剂量参数以避免管电流饱和,否则剂量节省效果会显著减小。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare radiation doses between conventional and chest pain protocols using dual-source retrospectively electrocardiography (ECG)-gated cardiothoracic computed tomography (CT) in children and adults and assess the effect of tube current saturation on radiation dose reduction.

MATERIALS AND METHODS: This study included 104 patients (16.6 ± 7.7 years, range 5-48 years) that were divided into two groups: those with and those without tube current saturation. The estimated radiation doses of retrospectively ECG-gated spiral cardiothoracic CT were compared between conventional, uniphasic, and biphasic chest pain protocols acquired with the same imaging parameters in the same patients by using paired t tests. Dose reduction percentages, patient ages, volume CT dose index values, and tube current time products per rotation were compared between the two groups by using unpaired t tests. A p value < 0.05 was considered significant.

RESULTS: The volume CT dose index values of the biphasic chest pain protocol (10.8 ± 3.9 mGy) were significantly lower than those of the conventional protocol (12.2 ± 4.7 mGy, p < 0.001) and those of the uniphasic chest pain protocol (12.9 ± 4.9 mGy, p < 0.001). The dose-saving effect of biphasic chest pain protocol was significantly less with a saturated tube current (4.5 ± 10.2%) than with unsaturated tube current method (14.8 ± 11.5%, p < 0.001). In 76 patients using 100 kVp, patient age showed no significant differences between the groups with and without tube current saturation in all protocols (p > 0.05); the groups with tube current saturation showed significantly higher volume CT dose index values (p < 0.01) and tube current time product per rotation (p < 0.001) than the groups without tube current saturation in all protocols.

CONCLUSION: The radiation dose of dual-source retrospectively ECG-gated spiral cardiothoracic CT can be reduced by approximately 15% by using the biphasic chest pain protocol instead of the conventional protocol in children and adults if radiation dose parameters are further optimized to avoid tube current saturation.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Goo HW. Comparison of Chest Pain Protocols for Electrocardiography-Gated Dual-Source Cardiothoracic CT in Children and Adults: The Effect of Tube Current Saturation on Radiation Dose Reduction. Korean J Radiol. 2018. PMID: 29353996. doi:10.3348/kjr.2018.19.1.23

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。