具有壁硬度与厚度区域变异的动脉体模
研究开发了一种多段动脉体模设计框架,基于多部件注塑成型原理,用聚乙烯醇冷冻凝胶依次浇铸病变段与侧翼段,制作出中央狭窄段(直径5.1 mm、壁厚1.95 mm、弹性模量212.6 kPa)与两侧翼段(直径6.0 mm、壁厚1.5 mm、弹性模量133.7 kPa)不同的三段体模,B型超声证实厚度差异,多普勒壁位移分析显示病变段峰值壁位移0.28 mm显著小于侧翼段0.60 mm。
要点
- 多部件注塑成型可制作壁硬度与厚度呈区域性差异的多段动脉体模。
- B型超声可确认病变段与侧翼段的壁厚差异。
- 搏动流下病变段峰值壁位移0.28 mm,显著小于侧翼段的0.60 mm。
核心结论
- 这种新的多段动脉体模有望作为评估局部动脉硬度估计算法的性能测试平台。
研究设计与样本
- 研究论文,开发基于多部件注塑成型原理的动脉体模设计框架,以聚乙烯醇冷冻凝胶为仿血管材料,制作并测试一个三段狭窄体模。
关键影像表现/征象
- B型超声确认体模病变段与侧翼段之间存在壁厚差异。
- 给予搏动流(颈动脉脉冲;峰值流速27 mL/s)时,病变段扩张程度小于侧翼段。
- 三搏动平均峰值壁位移在病变段为0.28 mm,显著小于侧翼段的0.60 mm。
诊断标准与关键数值
- 中央病变段:直径5.1 mm、壁厚1.95 mm、弹性模量212.6 kPa。
- 两侧翼段:直径6.0 mm、壁厚1.5 mm、弹性模量133.7 kPa。
- 多普勒壁位移分析条件:颈动脉脉冲、峰值流速27 mL/s、三搏动平均。
临床应用与阅片要点
- 该体模可用于评估局部动脉硬度估计算法,病变段与侧翼段的壁位移差异可作为算法性能测试的参考。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Regional wall stiffening and thickening are two common pathological features of arteries. To account for these two features, we developed a new arterial phantom design framework to facilitate the development of vessel models that contain a lesion segment whose wall stiffness and thickness differ from those of other segments. This new framework is based on multi-part injection molding principles that sequentially casted the lesion segment and the flank segments of the vessel model using molding parts devised with computer-aided design tools. The vessel-mimicking material is created from polyvinyl alcohol cryogel, and its acoustic properties are similar to those of arteries. As a case demonstration, we fabricated a stenosed three-segment phantom composed of a central lesion segment (5.1-mm diameter, 1.95-mm wall thickness, 212.6-kPa elastic modulus) and two flank segments (6.0-mm diameter, 1.5-mm wall thickness, 133.7-kPa elastic modulus). B-mode imaging confirmed the difference in thickness between the lesion segment and flank segments of the phantom. Also, Doppler-based vessel wall displacement analysis revealed that when pulsatile flow was fed through the phantom (carotid pulse; 27 mL/s peak flow rate), the lesion segment distended less compared with the flank segments. Specifically, the three-beat averaged peak wall displacement in the lesion segment was measured as 0.28 mm, and it was significantly smaller than that of the flank segments (0.60 mm). It is anticipated that this new multi-segment arterial phantom can serve as a performance testbed for the evaluation of local arterial stiffness estimation algorithms.
引用
- Chee AJY, Yiu BYS, Ho CK, et al. Arterial Phantoms with Regional Variations in Wall Stiffness and Thickness. Ultrasound Med Biol. 2018. PMID: 29361372. doi:10.1016/j.ultrasmedbio.2017.12.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。