MR成像靶向前列腺穿刺活检中的精确性:认知融合与弹性融合配准经直肠穿刺活检的前瞻性研究证据
这项前瞻性研究在88例前列腺影像报告和数据系统第1版评分≥3分的患者中,分别用认知融合和软件弹性融合配准技术各取2条穿刺组织,以穿刺针到靶灶中心及靶灶表面的距离衡量精确性;弹性融合配准的精确性更高(距离靶灶中心2.8 mm vs 7.1 mm,P<.0001),并在更多患者中检出癌症(40/44 vs 31/44,P=.03)。
要点
- 弹性融合配准穿刺精度高于认知融合(距离靶灶中心2.8 mm vs 7.1 mm,P<.0001)
- 弹性融合配准在更多患者中检出癌症(40/44 vs 31/44,P=.03)
- 穿刺阳性组织比阴性组织更靠近靶灶中心(1.7 mm vs 3.1 mm,P=.025)
核心结论
- 可变形MR成像/经直肠超声图像配准系统比认知徒手技术精确性更高,并在更多患者中检出癌症。
研究设计与样本
- 前瞻性研究,2013年6月至2013年9月进行,经机构审查委员会批准,患者提供口头知情同意;纳入88例前列腺影像报告和数据系统第1版评分至少3分的患者,分别用认知融合和弹性融合配准技术各取2条穿刺组织;精确性用欧几里得几何法测量,以活检的数字成像和通信医学档案计算穿刺组织到靶灶中心的距离(dCC)和到作为球体建模的靶灶表面的距离;采用Welch不等方差t检验和Yates校正χ2检验。
关键影像表现/征象
- 平均精确性为2.5 mm(95%CI:1.8 mm,3.3 mm)。
- 阳性穿刺组织比阴性穿刺组织更靠近靶灶中心(dCC:1.7 mm vs 3.1 mm,P=.025)。
- 靶内穿刺组织比靶外穿刺组织检出更多癌症(33/71条[46.5%] vs 3/17条[17.6%],P=.03)。
诊断标准与关键数值
- 精确性以dCC即穿刺组织到靶灶中心的距离和穿刺组织到作为球体建模的靶灶表面的距离衡量,精确性分析聚焦于每位患者或每种技术的最佳穿刺。
- 弹性融合配准技术的dCC为2.8 mm,认知融合技术为7.1 mm(P<.0001),弹性融合配准精确性更高。
- 认知融合技术在44例中检出31例癌症(70.5%),弹性融合配准在44例中检出40例(90.9%),P=.03。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To measure the precision in placement of a biopsy needle in a magnetic resonance (MR) imaging-detected target with transrectal ultrasonography (US), to document the clinical relevance of precision, and to report on the precision of cognitive and software-based registrations. Materials and Methods This prospective study was approved by the institutional review board and performed between June 2013 and September 2013. Patients provided informed verbal consent. Two cores each were obtained with cognitive and fusion techniques in 88 patients with a Prostate Imaging Reporting and Data System version 1 score of at least 3. Precision was measured with Euclidian geometry by using the Digital Imaging and Communications in Medicine archives of the biopsy as the distance from the core to the center (dCC) and the distance from the core to the surface of the target modeled as a sphere. To address clustering of data from multiple cores in the same patients, analyses of precision focused on the best shot for a patient or a technique. The Welch unequal variance t test and Yates corrected χ2 test were used as appropriate. Results Mean precision was 2.5 mm (95% confidence interval: 1.8 mm, 3.3 mm). Positive cores were closer to the center than were negative cores (dCC: 1.7 mm vs 3.1 mm, respectively; P = .025). More cancers were detected with on-target than off-target cores (33 of 71 cores [46.5%] vs three of 17 cores [17.6%]; P = .03). Cores obtained with the fusion technique achieved a higher precision than did cores obtained with the cognitive technique (dCC: 2.8 mm vs 7.1 mm, respectively; P < .0001). Targeted cores demonstrated cancer in 44 patients. Fewer cancers were detected with the cognitive technique than with the fusion technique (31 of 44 patients [70.5%] vs 40 of 44 patients [90.9%]; P = .03). Conclusion A deformable MR imaging/transrectal US image registration system achieved a higher precision and depicted cancer in more patients than did the cognitive freehand technique. © RSNA, 2018.
引用
- Cornud F, Roumiguié M, Barry de Longchamps N, et al. Precision Matters in MR Imaging-targeted Prostate Biopsies: Evidence from a Prospective Study of Cognitive and Elastic Fusion Registration Transrectal Biopsies. Radiology. 2018. PMID: 29361246. doi:10.1148/radiol.2017162916
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。